ESCO CCL-240B-8-IVF-HITEMP CelCulture® CO₂ 인큐베이터 (240 L, 230 VAC, 50/60 Hz)
생산업체: ESCO Model: CCL-240B-8-IVF-HITEMP P/N: 2170330 - 연락처
IR SENSOR MODEL WITH STAINLESS STEEL CHAMBER
CelCulture® Incubator 240L: IR Sensor, CO2 Control, ULPA, Moist Heat Decon, with Sealed Inner Door Kit for 240L (4/6 Glass Doors with Latches), Factory Installed, 230 VAC 50/60Hz
TEMPERATURE:
Temperature Control Method: Direct Heat and Air Jacket Using Microcontroller PI
Ambient Temperature Range*: 18 to 32°C (64.4 to 89.6°F)
Temperature Control Range, °C*: Ambient +5 to 60
Temperature Uniformity, °C*: <± 0.5
Temperature Fluctuation, °C*: <± 0.2
Temperature Recovery Time** (after 30 seconds door opening, 98% from initial value): ≤6 minutes
CO2:
CO2 Control System: Microcontroller PI
CO2 Control Range, % CO2: 0-19.5
CO2 Fluctuation, % CO2***: ± 0.2
CO2 Sensor: Infrared (IR) Sensor
CO2 Recovery Time**** (after 30 seconds door opening, 98% from initial value): Standard Unit: ≤5 minutes; Suppressed O2 model: ≤8 minutes
HUMIDITY:
Humidification Method: Humidity Pan
Humidity Range, %RH (at 37 °C)*******: 85-90
PHYSICAL CONSTRUCTION:
Interior Volume: 240 L
External Dimensions (W x D x H): 750 x 770 x 906 mm
Internal Dimensions (W x D x H): 595 x 640 x 633 mm
Chamber Construction: Main Body Electrogalvanized steel with Isocide™ antimicrobial coating
Interior Material: Stainless steel, type 304
Number of Shelves: 4
Maximum Number of Shelves: 7
Electrical Configuration 220-240 VAC, 50/60 Hz
Full Load Amps: 6.5 A
Net Weight: 121 kg
CONTAMINATION CONTROL
Contamination Control Methods
1) Main body is electro-galvanized steel with Isocide™ antimicrobial coating;
2) 90˚C moist heat OVERNIGHT decontamination cycle (HPA validated);
3) ULPA filter
4) 0.2-micron inlet filter for gas inputs;
5) 1-micron air circulation filter
All data recorded is specified for standard models with unloaded chambers and tested under optimum factory setting conditions of 23°C and 60% ambient humidity.
*Results are achieved when tested at 37°C as set point. Results may vary if set point changes and calibration is needed.
**For temperature not exceeding 37.2°C.
***Results are achieved when tested at 5% CO2 as set point. Results may vary if set point changes and calibration is needed.
****For CO2 level not exceeding 5.2%.
*****Results are achieved when tested at 5% O2 as set point. Results may vary if set point changes and calibration is needed.
****** For O2 level not lower than 4.8%.
******* Esco does not guarantee condensation-free chamber at humidity level higher than 90%.
- 품질 보증