直接联系认证专家

Division value (d): 0.1μg
Maximum range (Max): 2.1g
Repeatability (5% loading)
+ Standard deviation: 0.15μg
+ Typical error: ±0.3μg
Repeatability (Close to 100% load)
+ Standard deviation: 0.25μg
+ Typical error: ±0.5μg
Linear error
+ Standard deviation: 0.15μg
+Typical error ±0.3μg
Position error (Test location according to OIMLR76 specification)
+ Test Weight: 2g
+ Standard deviation: 0.15μg
+ Typical error: ±0.3μg
Sensitivity drift (+10℃~+30℃): 5PPM
Maximum peeling range: Less than 100% of the maximum range
Accuracy class: Class I
Verification index value (e) (According to 2014/31/EU specification): 1μg
Typical stability time: 12s
Typical measurement time:15s
External calibration weight: 2g
Standard weight grade: E2(accordance with OIML R111-1)
Pan Size: ϕ30mm