Specifications
 | 
 
| Model | 
Range  | 
Resolution | 
Accuracy | 
Burst Pressure | 
 
| Fluke-700G01 | 
-10 to +10 inH2O  -20 mbar to 20 mbar | 
0.001 inH2O,  0.001 mbar | 
Positive pressure ± 0.1 % of range,  Vacuum ± 0.1 % of range | 
3 psi,  210 mbar | 
 
| Fluke-700G02 | 
-1 to +1 psi,  -70 mbar to 70 mbar | 
0.0001 psi,  0.001 mbar | 
5 psi,  350 mbar | 
 
| Fluke-700G04 | 
-14 to 15 psi,  -0.97 to 1 bar | 
0.001 psi,  0.0001 bar | 
Positive pressure ± 0.05 % FS  Vacuum ± 0.1 % FS | 
60 psi,  4 bar | 
 
| Fluke-700G05  | 
-14 to 30 psi,  -0.97 to 2 bar | 
0.001 psi,  0.0001 bar | 
120 psi,  8 bar | 
 
| Fluke-700G06  | 
-12 to 100 psi,  -0.83 to 6.9 bar | 
0.01 psi,  0.0001 bar | 
400 psi,  26 bar | 
 
| Fluke-700G27  | 
-12 to 300 psi,  -0.83 to 20 bar | 
0.01 psi,  0.001 bar | 
1200 psi,  80 bar | 
 
| Fluke-700G07 | 
-12 to 500 psi,  -0.83 to 34 bar | 
0.01 psi,  0.001 bar | 
2000 psi,  138 bar | 
 
| Fluke-700G08 | 
-14 to 1000 psi,  -0.97 to 69 bar | 
0.1 psi,  0.001 bar | 
4000 psi,  266 bar | 
 
| Fluke-700G10 | 
-14 to 2000 psi,  -.97 bar to 140 bar | 
0.1 psi,  0.01 bar | 
8000 psi,  550 bar | 
 
| Fluke-700G29 | 
-14 to 3000 psi,  -0.97 to 200 bar | 
0.1 psi,  0.01 bar | 
10000 psi,  690 bar | 
 
| Fluke-700G30 | 
-14 to 5000 psi,  -0.97 to 340 bar | 
0.1 psi,  0.01 bar | 
10000 psi,  690 bar | 
 
| Fluke-700G31 | 
-14 to 10000 psi,  -0.97 to 690 bar | 
1 psi,  0.01 bar | 
20000 psi,  1035 bar | 
 
| Fluke-700GA4 | 
0 to 15 psia,  0 to 1 bar absolute | 
0.001 psi,  0.0001 bar | 
± 0.05 % of range | 
60 psi,  4 bar | 
 
| Fluke-700GA5 | 
0 to 30 psia  0 to 2 bar absolute | 
0.001 psi,  0.0001 bar | 
120 psi,  8 bar | 
 
| Fluke-700GA6 | 
0 to 100 psia,  0 to 6.9 bar absolute | 
0.01 psi,  0.0001 bar | 
400 psi,  27 bar | 
 
| Fluke-700GA27 | 
0 to 300 psia,  0 to 20 bar absolute | 
0.01 psi,  0.001 bar | 
1200 psi,  80 bar | 
 
| Fluke-700RG05 | 
-14 to +30 psi,  -0.97 to +2 bar | 
0.001 psi,  0.0001 bar | 
Positive pressure,  ± 0.04 % of reading +.01 % of range,  Vacuum ± 0.05 % of range | 
90 psi,  6 bar | 
 
| Fluke-700RG06 | 
-12 to 100 psi,  -0.83 to 6.9 bar | 
0.01 psi,  0.0001 bar | 
Positive pressure,  ± 0.04 % of reading +.01 % of range,  Vacuum ± 0.1 % of range | 
400 psi,  27 bar | 
 
| Fluke-700RG07 | 
-12 to 500 psi,  -0.83 to 34 | 
0.01 psi,  0.001 bar | 
2000 psi,  138 bar | 
 
| Fluke-700RG08 | 
-14 to 1000 psi,  -0.97 to 69 bar | 
0.1 psi,  0.001 bar | 
4000 psi,  275 bar | 
 
| Fluke-700RG29 | 
-14 to 3000 psi,  -0.97 to 200 bar | 
0.1 psi,  0.01 bar | 
10000 psi,  690 bar | 
 
| Fluke-700RG30 | 
-14 to 5000 psi,  -0.97 to 340 bar | 
0.1 psi,  0.01 bar | 
15000 psi,  1035 bar | 
 
| Fluke-700RG31 | 
-14 to 10000 psi,  -0.97 to 690 bar | 
1 psi,  0.01 bar | 
20000 psi,  1380 bar | 
 
700R Ranges: Temperature Compensation 0 e°C to 50 °C (32 °F to 122 °F) to rated accuracy.  For temperatures from -10 °C to 0 °C and 50 °C to 55 °C, add .005 % FS/°C | 
 
| 
 Media compatibility 
- 10 inH2O, 1, 15, 30 psi: any clean dry non-corrosive gas
 
- 100, 300, 500, 1000 psi: any liquids or gases compatible with 316 stainless steel and above 1000 psi: any non-flammable, non-toxic, non-explosive, non-oxidizing liquid or gas compatible with 316 stainless steel.
 
 
 | 
 
 
  
Mechanical Specifications
 | 
 
| Dimensions (HxWxD) | 
12.7 cm x 11.4 cm x 3.7 cm (5 in x 4.5 in x 1.5 in) | 
 
| Pressure connection | 
¼ in NPT Male | 
 
| Housing/Weight | 
Cast ZNAL 0.56 kg (1.22 lb) with holster | 
 
| Display | 
5-1/2 digits, 16.53 mm (0.65 in) high 20-segment bar graph, 0 to 100 % | 
 
| Power | 
Three (3) AA alkaline batteries | 
 
| Battery life | 
1,500 hours without backlight (continuous on), 2,000 hours at slow sample rate | 
 
 
 | 
 
 
 
 
 
 
 | 
 
 
 | 
  |