| Specifications |  
| Model | Range | Resolution | Accuracy | Burst Pressure |  
| Fluke-700G01 | -10 to +10 inH2O -20 mbar to 20 mbar
 | 0.001 inH2O, 0.001 mbar
 | Positive pressure ± 0.1 % of range, Vacuum ± 0.1 % of range
 | 3 psi, 210 mbar
 |  
| Fluke-700G02 | -1 to +1 psi, -70 mbar to 70 mbar
 | 0.0001 psi, 0.001 mbar
 | 5 psi, 350 mbar
 |  
| Fluke-700G04 | -14 to 15 psi, -0.97 to 1 bar
 | 0.001 psi, 0.0001 bar
 | Positive pressure ± 0.05 % FS Vacuum ± 0.1 % FS
 | 60 psi, 4 bar
 |  
| Fluke-700G05 | -14 to 30 psi, -0.97 to 2 bar
 | 0.001 psi, 0.0001 bar
 | 120 psi, 8 bar
 |  
| Fluke-700G06 | -12 to 100 psi, -0.83 to 6.9 bar
 | 0.01 psi, 0.0001 bar
 | 400 psi, 26 bar
 |  
| Fluke-700G27 | -12 to 300 psi, -0.83 to 20 bar
 | 0.01 psi, 0.001 bar
 | 1200 psi, 80 bar
 |  
| Fluke-700G07 | -12 to 500 psi, -0.83 to 34 bar
 | 0.01 psi, 0.001 bar
 | 2000 psi, 138 bar
 |  
| Fluke-700G08 | -14 to 1000 psi, -0.97 to 69 bar
 | 0.1 psi, 0.001 bar
 | 4000 psi, 266 bar
 |  
| Fluke-700G10 | -14 to 2000 psi, -.97 bar to 140 bar
 | 0.1 psi, 0.01 bar
 | 8000 psi, 550 bar
 |  
| Fluke-700G29 | -14 to 3000 psi, -0.97 to 200 bar
 | 0.1 psi, 0.01 bar
 | 10000 psi, 690 bar
 |  
| Fluke-700G30 | -14 to 5000 psi, -0.97 to 340 bar
 | 0.1 psi, 0.01 bar
 | 10000 psi, 690 bar
 |  
| Fluke-700G31 | -14 to 10000 psi, -0.97 to 690 bar
 | 1 psi, 0.01 bar
 | 20000 psi, 1035 bar
 |  
| Fluke-700GA4 | 0 to 15 psia, 0 to 1 bar absolute
 | 0.001 psi, 0.0001 bar
 | ± 0.05 % of range | 60 psi, 4 bar
 |  
| Fluke-700GA5 | 0 to 30 psia 0 to 2 bar absolute
 | 0.001 psi, 0.0001 bar
 | 120 psi, 8 bar
 |  
| Fluke-700GA6 | 0 to 100 psia, 0 to 6.9 bar absolute
 | 0.01 psi, 0.0001 bar
 | 400 psi, 27 bar
 |  
| Fluke-700GA27 | 0 to 300 psia, 0 to 20 bar absolute
 | 0.01 psi, 0.001 bar
 | 1200 psi, 80 bar
 |  
| Fluke-700RG05 | -14 to +30 psi, -0.97 to +2 bar
 | 0.001 psi, 0.0001 bar
 | Positive pressure, ± 0.04 % of reading +.01 % of range,
 Vacuum ± 0.05 % of range
 | 90 psi, 6 bar
 |  
| Fluke-700RG06 | -12 to 100 psi, -0.83 to 6.9 bar
 | 0.01 psi, 0.0001 bar
 | Positive pressure, ± 0.04 % of reading +.01 % of range,
 Vacuum ± 0.1 % of range
 | 400 psi, 27 bar
 |  
| Fluke-700RG07 | -12 to 500 psi, -0.83 to 34
 | 0.01 psi, 0.001 bar
 | 2000 psi, 138 bar
 |  
| Fluke-700RG08 | -14 to 1000 psi, -0.97 to 69 bar
 | 0.1 psi, 0.001 bar
 | 4000 psi, 275 bar
 |  
| Fluke-700RG29 | -14 to 3000 psi, -0.97 to 200 bar
 | 0.1 psi, 0.01 bar
 | 10000 psi, 690 bar
 |  
| Fluke-700RG30 | -14 to 5000 psi, -0.97 to 340 bar
 | 0.1 psi, 0.01 bar
 | 15000 psi, 1035 bar
 |  
| Fluke-700RG31 | -14 to 10000 psi, -0.97 to 690 bar
 | 1 psi, 0.01 bar
 | 20000 psi, 1380 bar
 |  
| 700R Ranges: Temperature Compensation 0 e°C to 50 °C (32 °F to 122 °F) to rated accuracy. For temperatures from -10 °C to 0 °C and 50 °C to 55 °C, add .005 % FS/°C
 |  
| Media compatibility 
10 inH2O, 1, 15, 30 psi: any clean dry non-corrosive gas100, 300, 500, 1000 psi: any liquids or gases compatible with 316 stainless steel and above 1000 psi: any non-flammable, non-toxic, non-explosive, non-oxidizing liquid or gas compatible with 316 stainless steel. |    
| Mechanical Specifications |  
| Dimensions (HxWxD) | 12.7 cm x 11.4 cm x 3.7 cm (5 in x 4.5 in x 1.5 in) |  
| Pressure connection | ¼ in NPT Male |  
| Housing/Weight | Cast ZNAL 0.56 kg (1.22 lb) with holster |  
| Display | 5-1/2 digits, 16.53 mm (0.65 in) high 20-segment bar graph, 0 to 100 % |  
| Power | Three (3) AA alkaline batteries |  
| Battery life | 1,500 hours without backlight (continuous on), 2,000 hours at slow sample rate |  |  |  |  |