Gastec 6A 蒸汽探测管H2O (20 ~ 80 mg/m3)
制造商: Gastec Model: 6A - 报价要求
- 联系
Indicator: 20 ~ 80 mg/m3
Measuring Range: 20 ~ 80 mg/m3
Sampling Rate: 100mL/min (500mL)
Sampling Time: 5 minute
Colour Change: Yellow → Green
Reaction Principle: H2O + Mg(ClO4)2 → Mg(ClO4)2 H2O
Coefficient of Variation: 10% (20 mg/m3),5% (for 20 to 80 mg/m3)
Nr. of tests: 10
Performance
Measuring Range | 20 to 80 mg/m3 |
Sampling Rate | 100mL/min (500mL) |
Correction Factor | 1 |
Sampling Time | 5 minutes |
Colour Change | Yellow → Green May produce "purple" colour at higher concentrations |
Reaction Principle | H2O + Mg(ClO4)2 → Mg(ClO4)2 H2O |
Coefficient of Variation | 10% (20 mg/m3),5% (for 20 to 80 mg/m3) |
Shelf Life | 2 Years |
Corrections for temperature | Unnecessary for 0 to 40℃ |
Store the tubes at cool and dark place. |
Calibration gas generation | Diffusion tube method |
Airtec tube - Compressed breathing air detector tubes
Measurement procedure:
1.Install a reducer pressure with gauge and flow metre to a cylinder, compressor or air line and adjust the flow metre the required setting. | ||
2.Break tips off a fresh detector tube in the tube tip breaker and insert a tube into a tube holder. | ||
3.Attach the rubber tube holder to the flow metre outlet. Make certain the tube arrowon the tube pointing in a down direction. | ||
4.Turn on the cylinder or compressor and confirm the flow metre according to each Airtec tube specifications. | ||
5.Time the flow rate with a stopwatch. | ||
6.As soon as sampling time has elapsed turn off the cylinder or compressor, and remove the tube from the tube holder and then read the colour change layer immediately. |