Gastec 6AH 蒸汽探测管 H2O
制造商: Gastec Model: 6AH 来源: Japan - 报价要求
- 联系
Indicator: 500 ~ 5000 ppm
Measuring Range: 500 ~ 5000 ppm
Sampling Rate: 300mL/min(300mL)
Sampling Time: 1 minute
Colour Change: Green → Purple
Reaction Principle: H2O + Mg(ClO4)2 → Mg(ClO4)2 H2O
Coefficient of Variation: 10% (for 500 to 1000 ppm), 5% (for 1000 to 5000 ppm)
Nr. of tests: 10
Measuring Range | 500 to 5000 ppm |
Sampling Rate | 300mL/min(300mL) |
Correction Factor | 1 |
Sampling Time | 1 minutes |
Detecting Limit | 300ppm |
Colour Change | Green → Purple |
Reaction Principle | H2O + Mg(ClO4)2 → Mg(ClO4)2 H2O |
Coefficient of Variation | 10% (for 500 to 1000 ppm), 5% (for 1000 to 5000 ppm) |
Shelf Life | 2 Years |
Corrections for temperature | Unnecessary for 0 to 40℃ |
Store the tubes at cool and dark place. |
Calibration gas generation | Bubbling method |
Airtec tube - Compressed breathing air detector tubes
Measurement procedure:
1.Install a reducer pressure with gauge and flow metre to a cylinder, compressor or air line and adjust the flow metre the required setting. | ||
2.Break tips off a fresh detector tube in the tube tip breaker and insert a tube into a tube holder. | ||
3.Attach the rubber tube holder to the flow metre outlet. Make certain the tube arrowon the tube pointing in a down direction. | ||
4.Turn on the cylinder or compressor and confirm the flow metre according to each Airtec tube specifications. | ||
5.Time the flow rate with a stopwatch. | ||
6.As soon as sampling time has elapsed turn off the cylinder or compressor, and remove the tube from the tube holder and then read the colour change layer immediately. |