For full functionality of this site it is necessary to enable JavaScript.

KYORITSU 4102A 接地电阻测试仪

CaseApplications

The metering:

+Soilresistance: 0~ 12Ω/ 0~ 120Ω/ 0~ 1200Ω

+GroundVoltage[50.60Hz]:0 ~30VAC

+Accuracy:

+Soilresistance:± 3%

+GroundVoltage:± 3%

Source:R6P(AA) (1.5V)× 6

Size:105 (L)×158 (W)×70 (D)mm

Weight:600gapprox.

Accessories:

  +7095A(Earthresistancetestleads) ×1SET

(red-20m,10myellow-, green-5m)

+8032(Auxiliaryearthspikes) ×1SET

+7127A(Simplifiedmeasurementprobe) ×1SET

+R6P(AA)× 6

+SoftCase

Details
The latest circuit design permits the instrument to operate with the minimum of influence from earth voltage and earth resistance of auxiliary earth spikes.
Dust and drip proof. (designed to IEC 529 IP54)
Earth resistance value can be read directly from the scale.
Designed to meet IEC 61010-1 safety standard.
Capable of measuring earth voltage.
Small and lightweight. Shock resistant new case material.
2mA measuring current permits earth resistance tests without tripping earth leakage current breakers in the circuit under test.
Lead wire connection to C and P terminals and proper auxiliary earth resistance can be checked by "OK" lamp. Lead wire connection to C and E terminals is good when "OK" lamp is illuminated.
4102A
CE
4102A Soft Case Model
4102A-H Hard Case Model
4102A Soft Case Model
4102A-H Hard Case Model
Measurement
Ranges
Earth Resistance: 0~12 Ω/0~120 Ω/0~1200 Ω
Earth Voltage [50,60Hz]: 0~30V AC
Accuracy Earth Resistance: ±3% of full scale
Earth Voltage: ±3% of full scale
Overload Protection Earth Resistance: 276V AC for 10 seconds
across 2 of the 3 terminals
Earth Voltage: 276V AC for 1 minute
Safety Standard IEC 61010-1 CAT.III 300V Pollution Degree 2, IEC 61557
Applicable Standards IEC 60529 IP54
Withstand Voltage 3700V AC for 1 minute
Power Source R6P (AA) (1.5V) × 6
Dimensions 105(L) × 158(W) × 70(D)mm
Weight 600g approx.
Accessories 7095A (Earth resistance test leads) × 1set
(red-20m, yellow-10m, green-5m)
8032 (Auxiliary earth spikes) × 1set
7127A (Simplified measurement probe) × 1set
R6P (AA) × 6
Neck strap
Instruction Manual
Carrying Case : 9084 (Soft case)
  : 9164 (Hard case)
Optional

7100 (Precision measurement cord set)

 

manual

datasheet

  • 质量承诺
  • 正品保修
  • 送货到家
  • 交易简单化

注册收新闻 - 获得优惠活动的机会