Ainuo AN9651TH(F)三相電気安全性総合テスター

お問い合わせ
安全なチェックアウト
高品質な対応
簡単な交換・返品
配送対応可能

1) Ground Bond Test
Output Current Setting Range: AC3A~30A 50Hz/60Hz,1A/step
Upper Limit of Resistance: 3A≤output current≤10A, 10~600mΩ; 10A
Ground Bond Measurement Range: 10~600mΩ                                                                                                                                                                                                                                                                  
Compensation Resistance Measurement Range: 0~200mΩ 1mΩ
Test Time Setting Range: 0.5~999.9s 0.1s/step
2) Insulation Resistance Test
Rated Output Load/Current: 1VA (1000V/1mA) 1mA
Output Voltage Range: DC:500V/1000V <1%
Insulation Resistance Range: 1.0~2000MΩ
Insulation Resistance Lower Limit Setting Range: 1.0~99.9 MΩ, 0.1MΩ/step, 100~2000 MΩ 1 MΩ/step
3) AC Hipot Test
Rated Output Load: 500VA (AC:5000V/100mA)
Output Voltage Range /Frequency/Adjustable Rate: 500~5000V, 50Hz/60Hz, <2%
Leakage Current Preset Alarm: Upper limit: 0.10~99.99mA; Lower limit: 0.00~99.99mA , 0.01mA/step
Leakage Current Measurement Range: 0.10~99.99 mA
4) DC Hipot Test
Rated Output Load: 60VA (AC:6000V/10mA)
Output Voltage Range /Frequency/Adjustable Rate: 500~6000V
Leakage Current Preset Alarm: Upper limit: 50~9999μA; Lower limit: 0~9999μA , 1μA    
Leakage Current Measurement Range: 50~9999μA 
5) Leakage Current Test
Leakage Current Range/Resolution: Upper limit: 50~9999μA 1μA
Compensation Current Setting Range/ Resolution: 0~500μA 1μA
Leakage Type: dynamic, static
Three Phase Dynamic Mode: A mode: ABC-AB-BC-CA
6) Power Test
Power Setting Range/Resolution
+ Upper limit, Single phase:20~6000W Three phase:60~20000W
+ Lower limit, Single phase:0~6000W Three phase:0~20000W 1W/step
Voltage Measurement Range and Accuracy: AC:50.0V~500.0V, 45~65Hz: ±(0.5% of reading +2counts)
Current Measurement Range and Accuracy: AC:0.30~30.00A, 45~65Hz: ±(0.5% of reading +2counts)
Active Power Measurement Range and Accuracy: 20.0~999.9W; 1.000~9.999kW; 10.00~20.00kW; 45~65Hz: ±(0.5% of reading +2counts) 
7) Low Voltage Start Test
Upper limit: 0.30 ~30.00A 
Lower limit: 0.00 ~30.00A0.01A
Voltage Measurement Range and Accuracy: AC:50.0V~500.0V, 45~65Hz: ±(0.5% of reading +2counts)
Current Measurement Range and Accuracy: AC:0.30~30.00A, 45~65Hz: ±(0.5% of reading +2counts) 
 Isolated Transformer Capacity: Three Phase:20kVA, Single Phase: 6kVA 
 DUT Type Setting: Single Phase, Three phase Three line, Three phase Four line 
Dimension (WxHxD): 765x1766x665
Gross weight: 260kg

 

Product Detail

Sigle-phase/Up to 7 functions
Double load for paralleled testing; 7 functions in 1 electrical safety and properties’ test; double channel display together; electric ARC, load open-circuit and short-circuit test function.
Applicable to such small home appliances high-speed product lines.

Technical characteristics:
◆ Compliant with GB4706.1-2005, IEC60335-1:2004, UL60335-1:2002, GB4943.1-2001, standard requirements.
◆ Three phase,single phase load compatible; Three phase rated load capacity 20kVA, can custom 60kVA.
◆ Seven functions into one device: ground bond, insulation resistance, AC hipot, DC hipot, leakage current, power test and low voltage start test.
◆ 15-group / 8-step test condition, intelligent judgment, auto compensation, auto protection;High speed test, the fastest speed can be 0.5s.
◆ Voltage ramp-up and ramp-down, discharge function;Dynamic and static leakage current, hipot test.
◆ Auto compensation, auto protection, auto memory, auto judgment, voice and light alarm.
◆ Serial communication, print, foot pedal interface, outer alarm light, PLC, network card interface.
◆ For safety test, the precision is 2%, for power test, the precision is 5%;For DUT, can support both grounding and suspending.
◆ Function preview, single screen setting, no need flip, simple operation.

◆ Support both 50Hz and 60Hz frequency, Chinese and English operation interface.

Model

AN9651TH (F)(Three Phase/single phase7 in1)

Ground Bond Test

Empty Load Output Voltage

<12v span="">

Output Current Setting Range

AC3A~30A 50Hz/60Hz,1A/step

Output Current Setting Accuracy

±(2% of setting+0.2A)

Upper Limit of Resistance

3A≤output current≤10A, 10~600mΩ; 10A

Ground Bond Resolution/ Measurement Range/ Accuracy

1mΩ, 10~600mΩ, ±(2% of reading +2mΩ)

Compensation Resistance Measurement Range/Accuracy

0~200mΩ 1mΩ

Test Time Setting Range

0.5~999.9s 0.1s/step

Insulation Resistance Test

Rated Output Load/Current

1VA (1000V/1mA) 1mA

Output Voltage Range/Adjustable

DC:500V/1000V <1% (Empty load to full load)

Output Voltage Ripple

Pure resistance load, 1000V/1mA, <3%

Insulation Resistance Range

1.0~2000MΩ

Insulation Resistance Lower Limit Setting Range/ Resolution

1.0~99.9 MΩ, 0.1MΩ/step, 100~2000 MΩ 1 MΩ/step

Insulation Resistance Accuracy

1~200 MΩ: ±(2% of reading+2counts) , 200~2000 MΩ: ±(5% of reading+5counts)

Ramp-up Time Range/Resolution

0.1~999.9s 0.1s

Delay Judgment Time

0.3~10.0s (Count from ramp-up )

Test Time Setting Range

0.5~999.9s 0.1s/step

AC Hipot Test

Rated Output Load

500VA (AC:5000V/100mA)

Output Voltage Range /Frequency/Adjustable Rate

500~5000V, 50Hz/60Hz, <2%(Empty load to full load)

Output Voltage Setting Range/Accuracy

Cold state AC:500~5000V, Hot state AC:500~2500V, 5V/step, ±(2.5% of setting+5V)

Leakage Current Preset Alarm

Upper limit: 0.10~99.99mA; Lower limit: 0.00~99.99mA , 0.01mA/step

Compensation Current Setting Range/ Resolution

0.00 ~5.00mA, 0.01mA

Output Initial Voltage Range

(0%~99%) of test voltage, from 0V to initial voltage need 200ms

Leakage Current Measurement Range/Resolution/Accuracy

0.10~99.99 mA, 0.01mA, ±(2% of reading +2counts)

Test Time/Ramp-Up/Ramp-Down

0.5~999.9s, 0.1~999.9s, 0.0~999.9s(After test pass)

DC Hipot Test

Rated Output Load

60VA (AC:6000V/10mA)

Output Voltage Range /Frequency/Adjustable Rate

500~6000V, <2%(Empty load to full load)

Output Voltage Setting Range/Accuracy

Cold state AC:500~6000V, Hot state AC:500~2500V, 5V/step, ±(2.5% of setting+5V)

Leakage Current Preset Alarm

Upper limit: 50~9999μA; Lower limit: 0~9999μA , 1μA

Compensation Current Setting Range/ Resolution

0 ~5000μA, 1μA

Output Initial Voltage Range

(0~99%) of test voltage, from 0V to initial voltage need 200ms

Leakage Current Measurement Range/Resolution/Accuracy

50~9999μA , 1μA, ±(2% of reading +2counts)

Test Time/Ramp-Up/Ramp-Down

0.5~999.9s, 0.1~999.9s, 0.3~10.0s(Count from ramp-up)

Leakage Current Test

Output Voltage Range

1.06 times of input voltage (output by isolated transformer)

Leakage Current Range/Resolution

Upper limit: 50~9999μA 1μA

Compensation Current Setting Range/ Resolution

0~500μA 1μA

Test Time Range/Resolution

1.0 ~999.9s 0.1s

Leakage Type

dynamic, static

Three Phase Dynamic Mode

A mode: ABC-AB-BC-CA (standard)
B mode: ABC-AB-ABC-BC-ABC-CA (customized)

Output Voltage Measurement Range/Accuracy

AC 50~500.0V 0.1V, ±(0.5% of reading+2counts)

Leakage Current Measurement Range/Accuracy

50~9999μA 1μA
DC~10kHz: ±(2% of reading+5counts) 10kHz~1MHz: ±(5% of reading+5counts)

Leakage Current Accuracy

50.0~2000μA is ±(5% of reading);2000μA or above is ±(2.5% of reading)

Leakage Power

6kVA ,isolated transformer capacity is designed according to the rated voltage 220V

Time Setting Range

1~300s

Measurement Mode

dynamic, static

Power Test

Output Voltage Range

1.00 times of input voltage (output by isolated transformer)

Power Setting Range/Resolution

Upper limit, Single phase:20~6000W Three phase:60~20000W
Lower limit, Single phase:0~6000W Three phase:0~20000W 1W/step

Test Time Range/Resolution

0.5 ~999.9s 0.1s

Voltage Measurement Range and Accuracy

AC:50.0V~500.0V, 45~65Hz: ±(0.5% of reading +2counts)

Current Measurement Range and Accuracy

AC:0.30~30.00A, 45~65Hz: ±(0.5% of reading +2counts)

Active Power Measurement Range and Accuracy

20.0~999.9W; 1.000~9.999kW; 10.00~20.00kW
45~65Hz: ±(0.5% of reading +2counts)

Power Factor Range and Accuracy

0.100~1.000, ±(2% of reading +2counts)

Low Voltage Start Test

Output Voltage Range

0.85 times of input voltage (output by isolated transformer)

Test Time Range/Resolution

0.5 ~999.9s 0.1s

Current Setting Range/Resolution

Upper limit: 0.30 ~30.00A Lower limit: 0.00 ~30.00A0.01A

Voltage Measurement Range and Accuracy

AC:50.0V~500.0V, 45~65Hz: ±(0.5% of reading +2counts)

Current Measurement Range and Accuracy

AC:0.30~30.00A, 45~65Hz: ±(0.5% of reading +2counts)

Display Mode

10.4 Inch, 640×480, TFT LCD display

Isolated Transformer Capacity

Three Phase:20kVA, Single Phase: 6kVA

Over Current Protection

Power, Start, Leakage Current, Phase Current35A, over 5s protection

DUT Type Setting

Single Phase, Three phase Three line, Three phase Four line

<span style="font-size:9.0pt; font-f