TECHNICAL SPECIFICATIONS

ORP Measure

Principle of ORP measure: Combination Electrode (ORP/reference) platinium ring,

                                            Reference Ag/AgCI. Gelled electrolyte (KCI)

Range of measures: - 1000.0 to + 1000.0 mV

Resolution: +/- 0.1 mV

Precision: +/- 10 mV

Answer time: < 90 s

Temperature measurement

Principle of measure T°C: CTN

Temperature: 0,00 °C to + 50.00°C

Resolution: 0.01 °C

Precision: +/- 0.5 °C

T90: < 300 s

Temperature of storage: 0°C to + 60°C

Protection scale: IP 68

Interface signal: Modbus RS-485 standard and SDI-12

Refresh rate measurement: Maximum < 1 second

Sensor power: 5 to 12 volts

Consumption: Standby : 25 μA

                       Average RS485 (1 measure/second) : 20 Ma

                      Pulse current : 500 mA

                      Heating time : 100 mS

Sensor

Dimensions of sensor mounted: Mounted sensor length: gland (262 mm) not included ;

                                                   Length with gland : 324 mm.

Weight: 350 g (sensor + cable)

Material in contact with the environment: PVC, POM-C,platinium, Polyurethane

Maximum Pression: 5 bars

Cable/ connection: 9 armored connectors, polyurethane jacket, bare wires or waterproof metal Fischer connector