Column Oven: 

Temperature range: Ambient temperature +7℃ ~ 400℃  (in 1℃ increment)

Temperature accuracy:  ± 0.02℃

Cooling time: 400℃ to 50℃ in 8-10 min at 25℃ ambient

Programmed temperature setting: 0.1℃~40℃/min (in 1℃ increment)

Program ramps: 7 ramps in total (10 ramps available with control workstation)

Size (L*W*H): 284*280*241mm(inside); 340*345*281mm(outside)

Flame Ionization Detector (FID): 

Detection limit: ≤3×10-12g/s (C16)

Best test result: ≤3×10-12g/s (C16)

Drift: ≤6×10-13A/30min

Noise: ≤5×10-14A

Linearity range: ≥106

Thermal Conductivity Detector (TCD): 

Sensitivity: ≥5000mV, mL/mg (C16)

Baseline noise: ≤20µV

Drift: ≤60µV/h

Linearity range: ≥104

Flame Photometric Detector (FPD): 

Detection limit: ≤8×10-13g/s(P) ; ≤8×10-11g/s(S)

Drift: ≤2×10-11A/30min

Baseline noise: ≤5×10-12A

Nitrogen-Phosphorus Detector (NPD): 

Detection limit: ≤5×10-12g/s(N) (Azobenzene); ≤5×10-13g/s(P) (Malathion)