Fluke (Biomedical) PS320 Fetal Simulator
Fetal ECG
Static rates: 0 BPM, 60 BPM, 90 BPM, 120 BPM, 150 BPM, 180 BPM, 210 BPM, and 240 BPM.
ECG sensitivity: 50 μV, 100 μV, 200 μV, 0.5 mV, 1 mV, and 2 mV.
US-1: Tracks primary fetal ECG rates.
US-2: Tracks secondary fetal activity for either independent “normal” or “twins” simulation; US-2 rate is fixed at 140 BPM.
Fetal patterns
Note: US-1 and fetal ECG track these selections. US-2 is in normal pattern, except during TREND #1 selection.
Trend #1: Twin fetal patterns
Normal: Normal pattern
Tachycardia: Tachycardia pattern
Bradycardia: Bradycardia pattern
Arrhythmias: Arrhythmia pattern
Late deceleration: Late deceleration
Early deceleration: Early deceleration
Moderate deceleration: Moderate variable deceleration
Acceleration #1: Acceleration wave #1
Acceleration #2: Acceleration wave #2
Sinusoidal (high): Sinusoidal pattern, large change
Sinusoidal (low): Sinusoidal pattern, small change
Severe variable deceleration #1: Severe deceleration wave #2
Severe variable deceleration #2: Severe variable deceleration
Prolonged deceleration: Prolonged deceleration
Biphasic deceleration: Biphasic deceleration
Exaggerated deceleration: Exaggerated deceleration
Non-uniform deceleration: Non-uniform deceleration
Variable deceleration (u): Variable deceleration, “U” shaped
Variable deceleration tach: Variable deceleration with high-rate BPM
Variable deceleration (v): Variable deceleration, “V” shaped
Variable deceleration (post): Variable deceleration, post exaggerated
Variable deceleration: Variable deceleration
Deceleration (position): Variable deceleration with position changes
Long deceleration: Long deceleration
Compensatory acceleration: Compensatory acceleration
Variability selections (added to fetal ECG)
Absent variability
Low variability
Mild variability
High variability
Severe variability
Long-term variability
Note: These patterns repeat and the TOCO channel will perform the selected TOCO waveform.
Maternal ECG
ECG static rates: 60 BPM, 80 BPM, 100 BPM, 120 BPM, 140 BPM, and 160 BPM.
ECG sensitivity: 0.5 mV, 1 mV, and 2 mV.
Pattern: Pattern selected during Trend #1 selection.
Uterine activity
Note: TOCO waveform selection is not available during Trend #1.
Execute waveform: Start TOCO waveform.
Uterine wave off: Stop TOCO waveform.
Analog: 0 V to 1 V; analog range 0 V to 1 V (1 V = 100 mmHg).
Uterine wave 0 to 25: Range of TOCO waveform.
Uterine wave 0 to 50: Range of TOCO waveform.
Uterine wave 0 to 100: Range of TOCO waveform.
Short duration: TOCO waveform of short duration.
Normal duration: Normal duration of TOCO waveform.
Increased duration: Long duration of TOCO waveform.
Uterine level = zero: Zero TOCO channel (automatic on power-up).
Uterine static +20: Increase TOCO static level by 20 mmHg (0–100 mmHg).
Increase resting tone: Resting tone increases.
Coupling: 2 close TOCO waves.
Tripling: 3 close TOCO waves.
Uterine pressure sensitivity: 5 μV or 40 μV on power-up.