Temperature range (@ 23°C ambient, 0.95 emissivity): –15°C to 120°C

Display accuracy: ±0.40°C at –15°C

                             ±0.40°C at 0°C

                             ±0.50°C at 50°C

                             ±0.50°C at 100°C

                             ±0.55°C at 120°C

Stability: ±0.10°C at –15°C

              ±0.05°C at 0°C

              ±0.10°C at 120°C

Uniformity (5.0 in dia of center of target): ±0.15°C at –15°C

                                                                  ±0.10°C at 0°C

                                                                  ±0.25°C at 120°C

Uniformity (2.0 in dia of center of target): ±0.10°C at -15°C

                                                                  ±0.10°C at 0°C

                                                                   ±0.20°C at 120°C

Heating time: 15 min: –15°C to 120°C

                       14 min: 23°C to 120°C

Cooling time: 15 min: 120°C to 23°C

                       20 min: 23°C to –15°C

Stabilization time: 10 minutes

Power: 115 V AC (±10%), 6.3 A, 50/60 Hz, 630 W

              230 V AC (±10%), 3.15 A, 50/60 Hz, 630 W

Fuse(s): 115 V AC 6.3 A, 250 V, slow blow

               230 V AC 3.15 A, 250 V, T

(H x W x D): 356 x 241 x 216 mm (14 x 9.5 x 8.5 in)

Weight: 9.1 kg (20 lb)