Flow Measurement

Flow ranges: ±3 cc/min, ±6 cc/min, ±30 cc/min, ±300 cc/min

Flow Units: 

±3.000 cc/min, ±180.0 cc/h, ±50.00 mm3/s, ±3.000 ml/min, ±180.0 ml/h

±6.000 cc/min, ±360.0 cc/h, ±100.0 mm3/s, ±6.000 ml/min, ±360.0 ml/h

±30.00 cc/min, ±1800 cc/h, ±500.0 mm3/s, ±30.00 ml/min, ±1800 ml/h

±300.0 cc/min, ±18000 cc/h, ±5000 mm3/s, ±300.0 ml/min, ±18000 ml/h

Accuracy @ 20°C:  <1% R + 0.003 cc/min, <1% R + 0.006 cc/min, <1% R + 0.03 cc/min, <1% R + 0.3 cc/min

Temperature Coefficients:

Zero: Semi-automatic Zero facility

Span: < 0.1% per °C

Display: 8 character LCD

Long Term Drift (span): < 1% per year

Electrical

Power Requirements: 9 volt battery (PP3/6LR61/MN1604)

Battery Life: Up to 100 hours

Operating Temperature: 0 to 50°C

Pneumatic

Media Compatibility: Clean dry air or non corrosive gas. Viscosity corrected for air only.

Overload: 20 x Range

Pneumatic Connection: ⅛” BSP female input (option for Quick Connector)

Vent port: M5 (DO NOT RESTRICT)

Static Pressure: Vacuum to 14 bar gauge

Effective Volume: 0.8cc (input port up to control valve)

Construction

Enclosure: ABS Case

Dimensions: 222 x 205 x 79mm (W x H x D)

Media Compatibility: Clean dry non-corrosive gas

Weight: 1.2kg