Gastec 133M 速い検出チューブテトラクロロエチレンCl2C:CCl2(2〜250ppm)
メーカーGastec(このブランドの他の製品を見る)
Model133M
お問い合わせ
安全なチェックアウト
高品質な対応
簡単な交換・返品
配送対応可能
Indicator: 5 to 100 ppm
Measuring Range:
+ 2 to 5 ppm, Number of Pump Strokes 2
+ 5 to 100 ppm, Number of Pump Strokes 1
+ 100 to 250 ppm, Number of Pump Strokes 1/2
Sampling Time: 45 seconds per pump stroke
Colour Change: Yellow → Reddish purple
Reaction Principle: Cl2C:CCl2 + PbO2 + H2SO4 → HCl, HCl + Base → Chloride
Coefficient of Variation: 10% (for 5 to 20 ppm), 5% (for 20 to 100 ppm)
Nr. of tests: 10
Pump: GV-100S/110S
Performance
| Measuring Range |
|
|||
| Number of Pump Strokes |
|
|||
| Correction Factor |
|
|||
| Sampling Time | 45 seconds per pump stroke | |||
| Detecting Limit | 0.4 ppm (n=2) | |||
| Colour Change | Yellow → Reddish purple | |||
| Reaction Principle | Cl2C:CCl2 + PbO2 + H2SO4 → HCl HCl + Base → Chloride |
|||
| Coefficient of Variation | 10% (for 5 to 20 ppm), 5% (for 20 to 100 ppm) | |||
| Shelf Life | 2 Years | |||
| Corrections for temperature & humidity | Temperature correction is necessary | |||
| Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below. | ||||
Possible coexisting substances and their interferences (NOTE)
| Substance | Concentration | Interference | Change colour by itself to |
| Nitric oxide, Nitrogen dioxide | No | No | |
| Chlorine, Bromine, Hydrogen chloride |
+ | Reddish purple | |
| Acetone | ≦ 200 ppm | No | No |
| Unsaturated halogenated HCs | + | Reddish purple | |
| Aromatic hydrocarbons | ≧ 100 ppm | No |
| Calibration gas generation | Diffusion tube method |
| TLV-TWA | TLV-STEL | Explosive range |
| 25 ppm | 100 ppm | - |
お得な情報を受け取る
数量割引、まとめ買い価格の更新、新製品情報をメールでお届けします。
登録することで、当社の利用規約およびプライバシーポリシーに同意したものとみなされます。
クイックサポート
認定専門家へ直接アクセス