For full functionality of this site it is necessary to enable JavaScript.
EMIN.VN
0
Product image

Gastec 154 Quick-measuring Detector tubes Cyclohexanone (C6H10O 154 (2~75ppm))

お問い合わせ
安全なチェックアウト
高品質な対応
簡単な交換・返品
配送対応可能

Indicator: 2 to 30 ppm
Measuring Range: 
+ 2 to 30 ppm, Number of Pump Strokes 4
+ 30 to 75 ppm, Number of Pump Strokes 2
Sampling Time: 3 minutes per pump stroke
Colour Change: Pale Yellow → Yellow
Coefficient of Variation: 10% (for 2 to 10 ppm), 5% (for 10 to 30 ppm)
Nr. of tests: 10
Pump: GV-100S/110S

Catalog

Manual

Performance
Measuring Range
2 to 30 ppm 30 to 75 ppm
Number of Pump Strokes
4 2
Correction Factor
1 2.5
Sampling Time 3 minutes per pump stroke
Detecting Limit 0.2 ppm (n=4)
Colour Change Pale Yellow → Yellow
Reaction Principle Cyclohexanone reacts with 2,4-dinitrophenylhydrazine to form dinitropnenylhydrazone to produce yellow in colour.
Coefficient of Variation 10% (for 2 to 10 ppm), 5% (for 10 to 30 ppm)
Shelf Life 2 Years
Corrections for temperature & humidity Temperature correction is necessary
Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Ketones   + Yellow
Other substance measurable with this detector tube
Substance Correction Factor No. of pump strokes Measuring Range
Furfural 1.0 4 2 to 30 ppm
Diacetone alcohol by scale 2 2.5 to 100 ppm
Isophorone 1.0 8 2 to 30 ppm
Calibration gas generation Diffusion tube method
TLV-TWA TLV-STEL Explosive range
20ppm(2014) 50ppm(2014) 1.1 to 9.4%