Indicator: 1 ~ 100 ppm
Measuring Range: 1 ~ 100 ppm
Sampling Time:  2 minute
Colour Change: White → Brown
Reaction Principle: S2- + Pb(CH3COO)2 → PbS
Nr. of tests: 10
Pump: GV-100

Catalog

Performance
Measuring Range 1 to 100 ppm
Sampling Time 2 minutes
Detecting Limit 0.5 ppm
Colour Change White → Brown
Reaction Principle S2- + Pb(CH3COO)2 → PbS
Shelf Life 3 Years
Corrections for temperature & humidity Unnecessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Changes colour by itself to
CO32- 500 ppm + No
Cl- ≧ 5000 ppm + No
CrO42- 25 ppm Yellow
SO32- 300 ppm No
S2O32- 50 ppm Unclear demarcation No

 

Detector tubes for dissolved substances
(i.e. in solution).
Dissolved sulphide detector tube
211H / 211M / 211 / 211LL
Liquid detector tube
 
Ozone: 218
Chloride ion: 221L / 221LL
Mercury: 271
Hexavalent chromium: 273
Bivalent iron: 281
Copper: 284
Zinc: 285
Nicke: 291
To measure, simply immerse the detector tube in a sample of liquid (placed in a beaker or other container). Anyone can measure concentrations quickly, easily and accurately on site.

* Adjustment for pH is required when measuring with liquid detector tubes.