For full functionality of this site it is necessary to enable JavaScript.
EMIN.VN
0
Product image

GEO Calibration 2000 SP-SH Humidity Generator (80 % RH)

お問い合わせ
安全なチェックアウト
高品質な対応
簡単な交換・返品
配送対応可能

GENERAL SPECIFICATIONS


OPERATING AMBIENT CONDITIONS

TEMPERATURE: 18 °C to 28 °C

HUMIDITY: Up to 80 % RH

STORAGE CONDITIONS

TEMPERATURE : -20 °C to 50 °C

HUMIDITY:  0 % to 95 % RH (non-condensing)

ALTITUDE : 2000 m

CALIBRATION SYSTEM ACCURACY

TEMPERATURE:  ± 0.10 °C Or Better Typically ± 0.05 °C Based on Probe Accuracy

HUMIDITY : ± 1.00 % RH Or Better Specially tuned systems can be as good as ± 0.60 %  Based on Probe Accuracy

CONTROLLER TYPE : PID Controller

UNIFORMITY

TEMPERATURE : 0.10 °C Or Better Typically ± 0.05 °C

HUMIDITY:  0.30 % RH @ 18 °C to 28 °C Or Better Typically ± 0.25 %

STABILITY

TEMPERATURE : 0.05 °C

HUMIDITY : 0.15 % RH @ 18 °C to 28 °C

WARM-UP PERIOD

AMBIENT CONDITIONS : 30 Minutes Maximum

COLD :30 Minutes

RESOLUTION

DISPLAY

Temperature : 0.01 °C

Humidity:  0.01 % RH

Dew Point : 0.01 °C (Calculated)

USB

Temperature 0.01 °C

Humidity 0.01 % RH

RAMP / SOAK RATE OF CHANGE

Depends on ambient and desiccant conditions

TEMPERATURE 

High → Low 1.50 °C / minute Typical

Low → High 5.00 °C / minute Typical

HUMIDITY 

High → Low 5.00 % RH / minute Typical

Low → High10.00 % RH / minute Typical

CONTROL RANGE

TEMPERATURE 5 °C to 55 °C

HUMIDITY@ 18°C 5 % to 95 % RH HUMIDITY@ 35°C 5 % to 80 % RH

HUMIDITY@ 23°C 5 % to 95 % RH HUMIDITY@ 40°C 5 % to 75 % RH

HUMIDITY@ 30°C 5 % to 85 % RH HUMIDITY@ 50°C 5 % to 60 % RH

Datasheet


お得な情報を受け取る

数量割引、まとめ買い価格の更新、新製品情報をメールでお届けします。

登録することで、当社の利用規約およびプライバシーポリシーに同意したものとみなされます。

クイックサポート

認定専門家へ直接アクセス