Automatic Buffer Recognition: Yes
Backlight: Yes
Barometric Pressure Measurement: Automatic compensation of DO when using an LDO or LBOD probe
Calibration: Demal (1D/ 0.1D/ 0.01D); Molar (0.1M/ 0.01M/0.001M); NaCl (0.05%; 25µS/cm; 1000µS/cm; 18mS/cm); Standard sea water; User defined
Calibration Intervals/Alerts/Reminder: Off, selectable from 2 hours to 7 days
Communication: Integrated USB type A (for USB 2.0 flash memory device, printer, keyboard) and Integrated USB type B (for PC)
Compliance Certifications: CE.WEEE
Conductivity measurement: Yes
Conductivity Measurement Range: 0.01 µS/cm - 200.0 mS/cm
Conductivity resolution: 0.01 µS/cm - 0.1 mS/cm upon selected measuring range
Contents: Meter Only
Custom Calibration Standards: Yes
Data Export: Download via USB connection to PC or flash memory device. Automatically transfer entire data log or as readings are taken.
Data Memory: 500 records/FIFO
Data Storage: Automatic, GLP ISO compliant reading data stored with calibration details.
Dimensions: 859 x 175 x 235 mm (3.38 x 6.88 x 9.25 in.)
Display: Detailed mode/Large mode
Display Type: 440 x 160 pixel LCD with backlight illumination
DO Measurement Range: 0.1 - 20.0 mg/L (ppm) 1 - 200% saturation
DO Resolution 0.1
DO sensor calibration:
• 100% (water-saturated air (100%) calibration
• 100% with 0 (water-saturated air (100%) calibration with 0 point
• mg/L (calibration with a specified dissolved oxygen concentration (mg/L) solution)
• mg/L with 0 (calibration with a specified dissolved oxygen concentration (mg/L) solution with 0 point)
• Factory (calibration with the default LDO calibration)
Electrode stand: Optional
GLP Features: Date; Time; Sample ID; Operator ID
Instrument: Bench
IP Rating: IP54 (Resistant to spray of water; Dust-proof)
ISE Direct Measurement Range: Yes
ISE Electrode Calibration: 2 - 5 points
Languages: English, French, German, Italian, Spanish, Danish, Dutch, Polish, Portuguese, Turkish, Swedish, Czech, Russian
mV Measurement at Stable Reading: Yes
mV Measurement Range: -1500 - 1500
mV Resolution: 0.1 mV
Needed Cable Length: NA
Operating Error Messages: Clear text error messages displayed
Operating Humidity: 90% relative humidity (non-condensing)
Operating Interface: Soft Touch Keypad
Operating Temperature: 0 - 60 °C (32 - 140 °F)
ORP Electrode Calibration: Predefined ORP standards (including Zobell's solution)
Parameter: pH/Oxydo Reduction Potential (ORP)
Conductivity/Total Dissolved Solid (TDS)/Salinity/Resistivity
Dissolved Oxygen (DO)
Biochemical Oxygen Demand (BOD)
Ion Selective Electrode (ISE): Ammonia, Ammonium, Chloride, Fluoride, Nitrate, Sodium
PC Data Transfer Software: HQD Series Meter Data Transfer Utility
pH Buffer Sets Color-coded: 4.01, 7.00, 10.01 pH;
IUPAC: 1.679, 4.005, 7.000, 10.012, 12.45 pH
DIN: 1.09, 4.65, 9.23 pH
User-defined custom buffer sets
pH Electrode calibration 1 - 3 Calibration points
Calibration summary data logged and displayed
pH Measurement Range: 0 - 14 pH
pH Resolution: Selectable: 0.001/0.01/0.1 pH
Printer: Yes, Optional
Probes included: None
Resistivity Measurement: 2.5 Ω.cm - 49 MΩ.cm
Resistivity Resolution: 0.1
Salinity Measurement Range: 0 - 42 (ppt) (‰)
Salinity Resolution: 0.01 (ppt) (‰)
TDS Measurement Range: 0.00 mg/L - 50.0 g/L as NaCl
TDS Resolution: 0.01 mg/L - 0.1 g/L upon measuring range.
Temperature Compensation: Automatic Temperature compensation for pH
Temperature Measurement: °C or °F
Temperature Range: -10 °C - 110 °C
Temperature Resolution: 0.1