PERFORMANCE

Range: 0 – 10^8 Pa.S

Measurement Time: 1 second

Accuracy: 1% of reading or +/- 1 digit

Repeatability: 0.3% of reading or +/- 1 digit

TEMPERATURE

Integral Thermometer: Integrated tip mounted temperature measurement

Range: 400 °C

PRESSURE

Standard Range: To process fitting

PROCESS CONNECTION

Standard fitting: All melt pipe fittings

Special fitting: Retrofits for all other viscometer replacements

ELECTRICAL CONNECTION

Cable Length: As required – not affected by cable length up to 1000m

Cable Type: Any standard instrument cable

Power: VP Processor powered

SAFETY CERTIFICATION

Hazardous Area: All instruments available for Ex Hazardous Area, explosive atmospheres

Ga Ex ia IIC T1…T6

CONSTRUCTION

All welded, crevice-free: No trap areas or dead-zones for polymer melt

MATERIALS

Head: SS316 ( EN 1.4401)

Fitting: As Required

Sensor: SS316 ( EN 1.4401) and SS316L ( EN 1.4404) as standard

Sensor/Fitting Options: Alloy C22 (EN 2.4602); Alloy C276 (EN 2.4819); S32750 Duplex Steel (EN 1.4410); Alloy 718

Special Materials: Non-standard materials to order *

SURFACE FINISH

Standard: N6 0.8 Ra

Special Finish: Non-standard finish options and coatings to order *

INGRESS PROTECTION

Standard: IP67

SOFTWARE

ViscoLink data viewing and logging PC software


お得な情報を受け取る