Rated voltage and effective measuring range: 125V-20MΩ,250V-50MΩ,500V-100MΩ

Center scale: 0.5MΩ/1MΩ/50MΩ

Minimum measurable resistance at rated voltage: 0.125MΩ,0.25MΩ,0.5MΩ

Rated current: 1mA+20%-0%

Open circuit voltage: Rated voltage+30%-0%

Short circuit current: <2mA

Withstanding Voltage: AC 3700V, 1 minute (Between input terminal and outer case)

Storage Temperature: -10℃~60℃, 80% RH max (Without condensation)

Operating Temperature: 0℃ to 40℃, 80% RH max. (Without condensation)

Power supply: 1.5V (AM-3, R6 or "AA" size)×6

Weight: Approx. 330g (Excluding batteries)

Accuracy (DC 50V): 

     First effective range:0.01MΩ~5MΩ ±5%rdg、10MΩ~50MΩ ±30%rdg

Accuracy(DC 125V): 

     First effective range:0.02MΩ~10MΩ ±5%rdg

     Second effective range:0.01MΩ~0.02MΩ 10MΩ~20MΩ ±10%rdg,10MΩ~100MΩ ±30%rdg

Accuracy (DC 250V): 

     First effective range:0.05MΩ~20MΩ ±5%rdg

     Second effective range:0.02MΩ~0.05MΩ 20MΩ~50MΩ ±10%rdg,50MΩ~100MΩ ±30%rdg

Accuracy (DC 500V): 

     First effective range:0.1MΩ~50MΩ ±5%rdg

     Second effective range:0.05MΩ~0.1MΩ 50MΩ~100MΩ ±10%rdg

Accuracy (DC 1000V): 

     First effective range:2MΩ~1000MΩ ±5%rdg

     Second effective range:1MΩ~2MΩ 1000MΩ~2000MΩ ±10%rdg

AC voltage (50/60Hz): 

     Range/Accuracy/Input impedance/Maximum input voltage

     AC 600V/±2.5% of full scale/Approx. 1.5M/AC 600V rms