For full functionality of this site it is necessary to enable JavaScript.
EMIN.VN
0
Product image

OMICRON P0000853 CT Analyzer

お問い合わせ
安全なチェックアウト
高品質な対応
簡単な交換・返品
配送対応可能

Accuracy

Ratio 1 ... 2000: error 0.02 % (typical) / 0.05 % (guaranteed)

Ratio 2000 ... 5000: error 0.03 % (typical) / 0.1 % (guaranteed)

Ratio 5000 ... 10000: error 0.05 % (typical) / 0.2 % (guaranteed)

Phase displacement

Resolution/; 0.01 min

Accuracy: 1 min (typical) / 3 min (guaranteed)

Winding resistance

Resolution: 1 mΩ

Accuracy: 0.05 % (typical) / 0.1 % + 1 mΩ (guaranteed)

Power supply

Input voltage: 100 VAC ... 240 VAC

Permissible input voltage: 85 VAC ...264 VAC

Frequency: 50 / 60 Hz

Permissible frequency: 45 Hz ... 65 Hz

Input power: 500 VA

Connection: Standard AC Socket IEC 60320

Output

Output voltage 0 ... 120 V

Output current 0 ... 5 Aeff (15 Apeak)

Output power 0 ... 400 VAeff (1500 VApeak)

Mechanical data

Size (W × H × D): 360 × 285 × 145 mm / 9.2 × 7.2 × 3.7 in

Weight: 8 kg / 17.4 lbs (without accessories)

Environmental conditions

Operating temperature: -10 °C ... + 50 °C / 14 °F ... 122 °F

Storage temperature: -25 °C ... + 70 °C / -13 °F ... 158 °F

Humidity: Relative humidity 5% ... 95% not condensing