Carbon Dioxide Sensor Type: Solid-state Infrared
Accuracy - Carbon Dioxide: Accuracy (for 0-50%, 0-100% range): ± 1% of reading, or ± 0.2% (whichever is greater), Accuracy (for 0-10%, 0-20% range): ± 1% of reading or ± 0.1% CO2 (whichever is greater, Accuracy (for 0-2000
CO2 Minimum Detection Limit: 0.1% (0-50% and 0-100% ranges), 0.01% (0-2.5%, 0-10% and 0-20% ranges), 10 ppm (0-5000ppm)
CO2 Calibration Controls: Potentiometer SPAN adjustment for CO2.
CO2 Calibration Frequency: Every 12 months; calibration gas recommended.
Expected Lifetime (oxygen): 4-5 years
Oxygen Range: 0.0 to 100%
Oxygen Sensitivity: 0.1% (or optional 0.01%)
O2 Resolution: 0.1% O2 (optional .01%)
Analog Output: 0-1Vdc linear proportional to concentration
Accuracy: : +/- 1% of reading; +/- 0.1% absolute below 10.0%
Drift: less than 0.4% per year
Oxygen Sensor Type: Proprietary Electrochemical
Response Time: 25 seconds to T95
O2 Sensor Lifetime: 4-5 Years
O2 Sensor Body: ABS Resin
Sample Connection: Front panel fitting with luer connection for disposable filter, plastic splitter; 1/16 to 1/8 ID tubing; (optional) Swaglok inlet
Sample Method: Pump (Optional)
Sample Pump: (Optional) Miniature diaphragm type with ~5cc/sec flow
Pump Timing: (Optional) 2-mode pump switch (timed, or continuous); (optional) cycling pump with on/off timer
Minimum Sample Required: 20cc
Sampling Assembly: Inlet fitting
Oxygen Calibration Controls: SPAN adjustment, zero adjustment
Instrument Calibration: Calibrates with room air; coarse SPAN adjustment on back; fine span adjustment on side; ZERO adjustment on side (calibration gas recommended for CO2)
Exhaust port: Vents to air OR tube outlet for recirculation to closed systems
Power Supply: 100-240V (50-60Hz) Charger included, with US or international interchangeable plug adapter (upon request)
Backup Power: Optional 12V internal battery pack
Battery Operation: Optional
Input Voltage: 100-240V, 50-60Hz
Battery Life: 8-12 hours before recharging
Measures: O2, CO2: O2, CO2
Form Factor: Benchtop
Size: 9.8W x 4H x 10L in. (249 x 102 x 254 mm)
Weight: 9.8 lb. (4.4 Kg)
Standards: CE, RoHS
Included Consumables: Plastic syringes for sample collection, 5 spare internal moisture blocking filters
Options: Battery, 0.01% resolution for O2
Included Documentation: Owners manual (includes operating instructions with detailed calibration instructions, calibration certficate, precautions), NIST calibration certificate
Origin of Goods: Our products are manufactured in the U.S.A.
Products by Gas: Oxygen and CO2
Products by Applications: Gas Flow / Process Analyzers