Carbon Dioxide Sensor Type: Solid-state Infrared
Accuracy - Carbon Dioxide: ± 1% of reading, or ± 0.2% CO2 (whichever is greater)
CO2 Minimum Detection Limit: 0.1% (0-100% range) or 0.01% (0-20% range)
CO2 Calibration Controls: Potentiometer SPAN adjustment for CO2 and zero adjusment
CO2 Calibration Frequency: Every 12 months; calibration gas recommended.
Expected Lifetime (oxygen): 4-5 years
Oxygen Range: 0.00 to 100%
Oxygen Sensitivity: 0.1% O2
O2 Resolution: 0.1% O2
Analog Output: Not Available
Accuracy: : +/- 1% of reading; +/- 0.1% absolute below 10.0%
Drift: less than 0.4% per year
Oxygen Sensor Type: Proprietary electrochemical cell, expected lifetime 4-5 years
Response Time: 15 seconds to T95
O2 Sensor Lifetime: 4-5 Years
O2 Sensor Body: ABS Resin
Sample Connection: Sample Probe
Sample Method: Pump
Sample Pump: Miniature diaphragm type with ~5cc/sec flow
Pump Timing: 2-12 second adjustable pump time, potentiometer on side of analyzer
Minimum Sample Required: 20cc
Sampling Assembly: Probe with female luer connection for filter and needle, low volume PTFE tubing
Oxygen Calibration Controls: SPAN adjustment, zero adjustment
Instrument Calibration: Calibrates with room air; coarse SPAN adjustment on back; fine span adjustment on side; ZERO adjustment on side (calibration gas recommended for CO2)
Exhaust port: Vents to air
Power Supply: Four "AA" NiMH rechargeable batteries; auto shut-off after 45 minutes idle. 100-240V (50-60Hz) Charger included, with US or international interchangeable plug adapter (upon request)
Backup Power: Four "AA" alkaline batteries (not included)
Battery Operation: Included
Input Voltage: 100-240V, 50-60Hz
Battery Life: 8-12 hours before recharging
Measures: O2, CO2: CO2
Form Factor: Portable
Size: 8 x 4 x 1.8 in. (200 x 100 x 40 mm)
Weight: 1 lb. (450 gms)
Standards: CE, RoHS
Included Consumables: 200 foam septas, 2 luer fit needles, 2 particulate filters, 2 liquid block filters
Included Documentation: Owners manual (includes operating instructions with detailed calibration instructions, calibration certficate, precautions), NIST calibration certificate
Origin of Goods: Our products are manufactured in the U.S.A.
Products by Gas: CO2
Products by Applications: Gas Flow / Process Analyzers