Scale Calibration Procedure

2021年04月26日 13時44分25秒

**1\. Range of application** This technical document specifies a calibration procedure for scale (non-automatic scales class I) with a division value not gre

1. Range of application

This technical document specifies a calibration procedure for scale (non-automatic scales class I) with a division value not greater than 1 mg. Technical scale is (non-automatic scale class II) according to OIML R76.

2. Calibrating techniques

- Outside check

- Technical check

- Measurement check

3. Calibrating instruments

3.1 Standard measuring instrument (s)

- Analytical Balance:
Kern ADB 200-4 Analytical balance (200g/0.1mg)
Sets of weights E2, Class E2, Measuring range: 1 mg ÷ 500g  
Sets of weights F1, Class F1, Measuring range: 1g ÷ 500g
- Technical Scale:

Sets of weights F1, Class F1, Measuring range 1g ÷ 500g
Standard Weight F1, Class F1, Measurement weight: 1 kg, 2 kg, 5 kg, 10 kg.
- Bench Scale:

Standard Weight M11, Class M1, Measuring range 0 ÷ 150kg

3.2 Auxiliary Instrument(s)
Hygro - Thermometer:

- Range of Measurement: (0 ÷ 50) °C
- Accuracy: ± 0,5 °C
- Range of Measurement: (10 ÷ 95) %RH
- Accuracy: ± 3 %RH

4. Conditions for Calibration
The calibration site must be bright, away from heat sources, wind sources, and not vibrated.
Temperature, humidity in the calibration site ensures:
Ambient temperature: (22 ± 3) ° C
Air humidity: (40 ÷ 80) %RH.
The scale should be calibrated at the site of use.

5. Prepare for Calibration
Before calibration, the following preparations must be done:
- The scale must be clean.
- Place the scale to be calibrated firmly on a surface and adjust the scale.
- Turn on the scale for at least 30 minutes or as required by the manufacturer.
- The scale to be calibrated should be thermally equilibrium with the environment in which it is calibrated. The minimum time required for this process is specified by the manufacturer.

6. Begin the calibration
6.1 Outside check
6.2 Technical Check
6.3 Measurement check
6.3.1 General requirements
6.3.2 Begin the calibration
- Test repeatability
- Check the deviation angle
- Check the correctness

7. Measurement Uncertainty Evaluation
8. General handling
- The calibrated scale is stamped, certified in paper.
- The calibration result includes standard value, indicator value, main number, together with the measurement uncertainty expanded for each weight level, coverage factor k.
- The recommended calibration cycle of scale is 12 months.

9. About EMIN's Calibration - Verification - Repair center
EMIN Vietnam’s Calibration and Verification center has been established since 2020 to provide after-sales services. The Calibration and Verification center is thoroughly equipped with technical equipment, followed by the assistance from various equipment manufacturers around the globe. At the center, calibration of specialized equipment is successfully carried out by experienced engineers in EMIN Vietnam, ensuring fast, accurate, reliable procedure to meet the most demanding customers’ requirements). Such technical capacity can only be performed by very few technology businesses in Vietnam. Meanwhile, after tireless effort, the Calibration and Verification center of EMIN Vietnam is officially certified to be accredited by the ISO 17025 standard (Jan, 2021)
If you want your instruments calibrated, please contact:
Equipment Verification - Calibration - Repair Service Center - EMIN Vietnam JSC 
Address: No 203 Nguyen Ngoc Vu Street, Trung Hoa Ward, Cau Giay District, Ha Noi City
Email: services@emin.vn - Hotline: 0968.060.220
Further information about scale calibration service can be found here.


関連ニュース

What Is Resonant Frequency? Why Can a Single Frequency Range Damage the Entire Product?
2026年07月10日 15時10分46秒

A device may operate stably for months, yet suffer cracks, broken solder joints, or snapped component leads after being transported. What surprises many engineers is that the vibration amplitude during transport is no greater than usual. The root cause lies in a specific frequency range where the product's structure absorbs energy most intensely, triggering resonance and causing continuous stress concentration at a single point.

Should Carton Packaging Undergo Vibration Testing or Drop Testing First?
2026年07月10日 10時38分40秒

Một kiện hàng gần như không bao giờ bị rơi ngay khi vừa rời khỏi nhà máy. Quãng đường hàng trăm kilomet trên xe tải, container hay băng chuyền mới là giai đoạn diễn ra đầu tiên, nơi bao bì liên tục chịu rung lắc trong nhiều giờ. Chỉ sau đó mới đến những cú va chạm khi bốc xếp, giao nhận hoặc rơi ngoài ý muốn.

Distinguishing Between DFT and WFT in Coating Thickness Measurement
2026年07月09日 14時40分30秒

Many projects encounter a situation where a freshly applied coating looks perfectly uniform and paint consumption matches the estimate, yet additional coating is required during inspection because the thickness is insufficient. This leads many to suspect measurement errors or equipment inaccuracy, whereas the actual cause is a confusion between WFT (Wet Film Thickness) and DFT (Dry Film Thickness)

Can a car paint thickness gauge detect if a vehicle has been in a collision?
2026年07月08日 16時56分01秒

Many used car buyers often spend a considerable amount of time examining the paintwork, the gaps between body panels, or the overall condition of the exterior. However, these signs can be skillfully masked following repairs; precise color matching and careful polishing can make the vehicle appear almost factory-original.

お得な情報を受け取る

数量割引、まとめ買い価格の更新、新製品情報をメールでお届けします。

登録することで、当社の利用規約およびプライバシーポリシーに同意したものとみなされます。

クイックサポート

認定専門家へ直接アクセス