For full functionality of this site it is necessary to enable JavaScript.
EMIN.VN
0
Product image

SE International PEN500mR Pen Dosimeters (Scale increments of 20mR)

ModelPEN500mR
原産地USA
お問い合わせ
安全なチェックアウト
高品質な対応
簡単な交換・返品
配送対応可能

Exposure Range (gamma and x-rays): Scale increments of 20mR

Energy Dependence: ±10% maximum change in sensitivity for x and gamma ray energies from 16 keV to 8 MeV.

Electrical Leakage

In the absence of radiation, the dosimeters leakage or self discharge is less than 0.5% of full scale in 24 hours at 50°C. At 20°C dosimeters will read background radiation. After gross over exposure totaling up to 2000 R the leakage is less than 5% of full scale in 48 hours at 20°C. Thereafter, the leakage will decrease to original values.

Charging Voltage

Dosimeters are capable of being charged to “zero” with 140 to 195 volts.

Operational Characteristics

A maximum change in sensitivity of ±10% may occur under any probable combination of the following:

A) Temperature range below -20°C (-4°F)

or above +50°C (122°F)

B) Greater than 50,000 foot altitude for one month

C) Relative humidity greater than 95%

Highly resistant to shock and vibration.

Accuracy: ±10% of true dose for Cs137 or Co60 gamma.

Rate Response: Dosimeters are dose rate independent.

Enclosure Material: Barrel - carbon fiber-filled, high-strength, electrically-conducting plastic with metal clip.

Weight: Approx. 19 grams (0.7 oz.)

Size

 Length: 11 cm (4.5 inches)

Diameter: 1.5 cm (0.6 inches)

Options

Protective sapphire lens

Datasheet

お得な情報を受け取る

数量割引、まとめ買い価格の更新、新製品情報をメールでお届けします。

登録することで、当社の利用規約およびプライバシーポリシーに同意したものとみなされます。