For full functionality of this site it is necessary to enable JavaScript.
EMIN.VN
0
Product image

TAITEC BR-40UL Constant Temperature Incubator Shaker (+4ºC ~ +70ºC; ±0.3 ~ 1.0℃)

ModelBR-40UL
P/N0090421-000
お問い合わせ
安全なチェックアウト
高品質な対応
簡単な交換・返品
配送対応可能

Temperature range (*1): +4ºC to +70ºC

Ambient temp. range (*1): +5℃ to +30℃

Temp. control accuracy (*2): ±0.3 to 1.0℃

Shaking motion: Reciprocal/Orbital

Speed range: 20 to 200 r/min

Shaking width: 10 to 40 mm(stepless variable type)

Door type: Swing lift-up door

Capacity:

Universal shaking platform MT-4430: Erlenmeyer flask 500mL × 8

Universal shaking platform MT-4030: Erlenmeyer flask 2L × 2

Heating method: Heater 500 W

Cooling method: Compressor 75 W, Cooling function can be switched on/off with the cooling switch.

Setting and display of temperature/speed: Digital setting, digital display

Platform size: 400 × 300 mm(max. inside height:331 mm)

Maximum load: approx. 7 kg (Platform and Vessel holder included)

Program functions: 1 program, 9 segments: Temp., Shaking speed and time/per segment (Setting range: 00 h 00 min to 99 h 59 min), Temp. gradient function

Safety devices/functions: Leakage/Overcurrent breaker, Fuse, High temp, Alarm of abnormal sample temp. (Set as desired),Detecting the door opening and closing, Self-diagnosis function for sensor error, Motor current limiter, Compressor overheating cut-off circuit

Other functions: Connecting hole inside chamber × 1

Dimensions (W × D × H): 586 × 636 × 689 Hmm

Weight: approx. 60 kg

Power supply: AC100V/9A

Standard accessories: Magnetic sheet × 2, Drain-reception plastic bottle with hose × 1

(*1)When using at the set temperature of +10.0°C or less, keep the environmental temperature at +30°C or less. (Reference: +4°C is available when the environmental temperature is +25°C.) The specifications of this product indicate the performance values according to the measurement methods specified by us.

(*2)Temperature change immediately after the defrosting function is activated is not included. In continuous operation at low temperature, frost may occur in the chamber and the above “Operating temperature range” and “Temperature control accuracy” may not be satisfied. Periodically defrost the chamber.