For full functionality of this site it is necessary to enable JavaScript.
EMIN.VN
0
Product image

TEKTRONIX EA-BIM 20005 バッテリーインピーダンスメーター (1 mHz ~ 10 kHz, 20CH)

ModelEA-BIM 20005
P/NEA-BIM 20005-10-20
お問い合わせ
安全なチェックアウト
高品質な対応
簡単な交換・返品
配送対応可能

AC Current Output (1 Hz - 10 kHz): Max. 10 A (peak-peak)

AC Current Output (100 mHz - 1 Hz): Max. 10 A (peak-peak)

AC Current Output (10 mHz - 100 mHz): Max. 5 A (peak-peak), depending on the cell voltage

AC Current Output (1 mHz - 10 mHz): Max. 2.5 A (peak-peak)

AC Excitation: Parallel multi-sine operation, up to 32 frequency points

DC Current Output (charge and discharge option): Max. ±1 A

BIM measurement charge/discharge - BIM measurement during temperature change: Yes. An impedance measurement can be performed in parallel during charging and discharging or temperature change tests, but care must be taken to select the lowest frequency and the excitation periods so that the state of the battery does not change significantly during the BIM measurement.

DC Voltage Range: 0 - 5 V

Frequency Range: 1 mHz - 10 kHz, the AC excitation frequency limits the AC excitation current. In the range of 1 Hz - 10 kHz the maximum current is limited to 10 A (peak-peak) and in frequencies below <1 Hz, the maximum current is limited to 5 A (peak-peak).

Frequency Accuracy: 100 ppm

Impedance Range: 0.1 mΩ to 100 mΩ

Interfaces: USB, CAN (free configurable)

Temperature Measurement Unit: 1 (PT100, 4-wire measurement)

Temperature channels: 20 (multiplexed)

Voltage Input Resolution (DC): < 0.5 mV

Voltage Input Resolution (AC): < 2 μV

Accuracy |Z| (typical, after calibration): < 1 %

Accuracy arg (Z) (typical, after calibration): < 0.5, depending on the cell voltage

Input Power: < 60 W

Input Voltage: 100 - 230 V AC

Dimensions (W x D x H): 19" x 19.7" x 5.25" (19" x 50 cm x 3U)

Ambient Temperature: +10°C to +30°C

Datasheet

お得な情報を受け取る

数量割引、まとめ買い価格の更新、新製品情報をメールでお届けします。

登録することで、当社の利用規約およびプライバシーポリシーに同意したものとみなされます。

クイックサポート

認定専門家へ直接アクセス