Vitrek HVL-35G DC High Voltage Probe (35 kVDC)
MAXIMUM VOLTAGE MEASUREMENT (INPUT IMPEDANCE)
Direct Input Terminal: 10KVDC, 10KVACrms (110 Meg ohms)
HVL-35G Probe: 35KVDC (10 Gig ohms)
DC VOLTAGE MEASUREMENT ACCURACY (RESOLUTION)
Direct Input Terminal: 0.03% of reading + 0.03V (10mV)
HVL-35G Probe: 0.25% of reading + 1.5V (1V)
GENERAL SPECIFICATIONS
AC/DC Voltage Measurement Resolution: Selectable 4, 5 or 6 digits
Measurement Functions: DC Voltage, True RMS AC Voltage, Ripple, Peak-to-Peak, Crest Factor & Frequency Measurement (4 digits, 0.01 Hz to 600 Hz)
Measurement Time: Selectable down to 16 mS (60 readings per sec.)
Maximum Input to Common Terminal: 3 KV Peak for 1 sec, no damage
Accuracy Specifications: Valid one year ±5°C from calibration temperature
Interfaces: LAN (Ethernet), RS232/Serial and USB printer port standard. GPIB optional.
ENVIRONMENTAL / PHYSICAL
Operating Environment: 0 to 50°C, <85% RH (non-condensing)
Dimensions: 218 mm (8.5″) wide × 130 mm (5″) high × 253 mm (10″) deep, nominally
Weight: 2.4 kg max (4 kg shipping)
POWER
AC Source: 45 to 450 Hz, 100 to 265 Vrms
DC Source: 11 to 16 VDC at <1.2 A, using a center-positive 2.5 mm DC power connector
BP-47 Battery Option: Up to 11 hours continuous operation