Fluke (Calibration) 1586A/2DS Super-DAQ 高精度温度スキャナー
生産者: Fluke (Calibration) Model: 1586A/2DS - 連絡先
Mains voltage:
+ 100 V setting: 90 V to 110 V
+ 120 V setting: 108 V to 132 V
+ 220 V setting: 198 V to 242 V
+ 240 V setting: 216 V to 264 V
Frequency: 47 Hz to 440 Hz
Power consumption: 36 VA peak (24 W average)
Environment temperature:
+ Operating: 0°C to 50°C
+ Full accuracy: 18°C to 28°C
+ Storage: −20°C to 70°C
+ Warm-up: 1 hour to full accuracy specifications
Relative humidity (non-condensing):
+ Operating: 0°C to 30°C <80%; 30°C to 50°C <50%
+ Storage: −20°C to 70°C <95%
Altitude
+ Operating: 2,000 m
+ Storage: 12,000 m
Vibration and shock: Complies with MIL-PRF-28800F Class 3
Channel capacity
+ Total analog channels: 45
+ Voltage/resistance channels: 41
+ Current channels: 5
+ Digital I/O: 8 bits
+ Totalizer: 1
+ Alarm outputs: 6
+ Trigger input: 1
Input protection: 50 V all functions, terminals and ranges
Math channels
+ Number of channels: 20
+ Operations: Sum, difference, multiply, divide, polynomial, power, square root, reciprocal, exponential, logarithm, absolute value, average, maximum, minimum
Triggers: interval, external (trigger input), alarm, remote (bus), manual, automated test
Memory
+ Scan data RAM: 75,000 readings with timestamp
+ Data/setup flash memory: 20 MB USB host port
+ Connector type: Type A
+ Function: Memory
+ File system: FAT32
+ Memory capacity: 32 GB
USB Device Port
+ Connector type: Type B
+ Class: Instrument
+ Function: Control and data transfer
+ Command protocol: SCPI
LAN
+ Function: Control and data transfer
+ Network protocols: Ethernet 10/100, TCP/IP
+ Command protocol: SCPI
RS-232
+ Connector: D-sub 9 pin (DE-9)
+ Baud rates: 1200, 2400, 4800, 9600, 19200, 38400
+ Function: Temperature source control output
(HxWxD): 150 mm x 245 mm x 385 mm
Weight: 6 kg (typical configuration)