For full functionality of this site it is necessary to enable JavaScript.

KONICA MINOLTA CA-P410 Display Color Analyzer (Normal Probe, 0.002 ~ 10,000 cd/m²)

Measurement area: Ø 10 mm

Acceptance angle: ± 5° 

Accuracy guaranteed measurement distance: 30 ± 5 mm

Luminance:

Accuracy guaranteed range: 0.002 to 10,000 cd/m2

Accuracy (for white):

> 0.0001 cd/m2: -

> 0.0003 cd/m2: -

> 0.0005 cd/m2: -

> 0.001 cd/m2: ± 9 % (0.002 to cd/m2)

> 0.01 cd/m2: ± 2.5 %

> 0.1 cd/m2: ± 2 %

> 1 cd/m2: ± 2 %

> 10 cd/m2: ± 1.5 %

> 100 cd/m2: ± 1.5 %

Chromaticity:

Accuracy guaranteed luminance range: 0.01 to 10,000 cd/m2

Accuracy (for white):

> 0.001 cd/m2: -

> 0.003 cd/m2: -

> 0.01 cd/m2: ± 0.006

> 0.1 cd/m2: ± 0.002

> 1 cd/m2: ± 0.002

> 10 cd/m2: ± 0.002

> 100 cd/m2: ± 0.002

Parameters:

Lvxy (∆Lv ∆x ∆y)

Lvu’v’(∆Lv ∆u’ ∆v’)

Lv TCP duv (∆Lv ∆TCP duv)

X Y Z (∆X ∆Y ∆Z)

Lv λd Pe (∆Lv ∆λd ∆Pe)

Flicker (Contrast)

Flicker (JEITA)

Measurement synchronization mode:

Measurement speed mode: AUTO, LTD. AUTO, SLOW, FAST

Measurement target (Vertical synchronization frequency): 0.5 to 240 Hz (luminance and chromaticity), 0.5 to 130 Hz (flicker)

User calibration memory channel: 99 channels

Interface:

Communication: USB2.0, RS-232C

Trigger: IN: 1.8 V /3.3 to 5 V switching Out: 5 V 

Power supply: DC 5 V (input from USB bus power line or RS communication connector)

Operation temperature/humidity range: 10 to 35°C, relative humidity 85% or less with no condensation

Storage temperature/humidity range: 0 to 45°C, relative humidity 85% or less (at 35°C) with no condensation