Memmert HPP750 一定気候チャンバー(749L)
生産者: MEMMERT Model: HPP750 起源: Germany 品物を保証します: 12 Month - 要件クオート
- 連絡先
Volume: 749 liters
Interior dimension: 1040W x 1200H x 600D mm
Exterior dimension: 1224W x 1726H x 856D mm
Stainless steel grids: 2
Max. number of grids: 14
Working temperature range without humidity: 0 (at least 20 below ambient temperature) to 70°C
Working temperature range with humidity: +5 (at least 20 below ambient temperature) to 70°C
Setting temperature range without humidity: 0 – 70°C
Setting temperature range with humidity: 5 – 70°C
Setting accuracy: 0.1°C
2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
Setting range humidity: 10 – 90% rh
Humidity supply with distilled water from external tank by self-priming pump
Humidification by hot steam generator
Dehumidification by cold trap using the Peltier technology
Setting accuracy: 0.5% rh
ControlCOCKPIT: TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays
Language settings: German, English, Spanish, French, Polish, Czech, Hungarian
Adjustable parameters: temperature (Celsius or Fahrenheit), relative humidity, programme time, time zones
Timer: Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function HeatBALANCE: adapting the distribution of the heating performance of the upper and lower heating circuit from -50 % to +50 %
Function SetpointWAIT: the process time does not start until the set temperature is reached.
Calibration: three freely selectable values each, temperature and humidity
Interface: Ethernet LAN, USB
Programme stored in case of power failure
Programming: AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port
ControlCOCKPIT with USB port for uploading programmes, reading out protocol logs, activating the User-ID function
Displaying of already logged protocol data on the ControlCOCKPIT (max 10000 values correspond to approx. 1 week)
Temperature control: over- and undertemperature monitor TWW, protection class 3.3 or adjustable temperature limiter TWB, protection class 2, selectable on display
Additionally integrated over and undertemperature protection "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over or undertemperature, heating function is switched off in case of overtemperature, cooling function in case of undertemperature
Integral fault diagnostics for temperature and humidity control
Alarm: visual and acoustic
Energy-saving Peltier heating-/cooling system integrated in the rear
Structured stainless steel housing, scratch-resistant, robust and durable; rear of zinc-plated steel
Door: fully insulated stainless steel door with 2-point locking, inner glass door, two leaves
Internal data logger with a storage capacity of least 10 years
Voltage: 230 V, 50/60 Hz
Electrical load: approx. 1400 W.
Ambient temperature: 16 – 40°C
Humidity: max. 70% rh, non-condensing
Installation: 4 feet; mounted on lockable castors
Accessories:
+ Calibration certificate standard values 10°C and 37°C, 60% rh at 30°C (Memmert supply)
+ Stainless steel grids, 2 pcs
+ Water tank including connection hose
+ Manual