For full functionality of this site it is necessary to enable JavaScript.
EMIN.VN
0
Product image

3ctest CSG 2050 Surge (Impulse) Current Simulator (20kA±10%)

문의하기
안전한 결제
품질 보장
간편 교환 및 반품
배송 가능

Output waveform: 20/50μs(±10%) current wave

Output peak current: 20kA±10%

Reversed polarity peak: <20%

DUT residual voltage: < 3kV

Voltage polarity: +/- (automatic switch)

Charging time: 10s~999s

Charging voltage instability: Less than 1%

Trigger mode: Automatic or manual, single trigger

Trigger malfunction: Less than 1%

Trigger range: 100% rated charging voltage

Measurement display resolution: Better than 3%

Insulation strength: ≤500V RMS(common port COMMCN to reference ground PE)

Auxiliary mains voltage: 220V(±10%)/50Hz

Auxiliary mains current: <10A(continuous)

- Calibration parameters:

Rise time: 20 μs±10%

Time to half value:  50μs±10%

First measurement range

Short circuit output peak current:

1kA-0%/﹢30% @2kV

2kA-0%/﹢30% @4kV

3kA-0%/﹢30% @6kV

4kA-0%/﹢30% @8kV

5kA-0%/﹢30% @10kV

Second measurement range

Short circuit output peak current:

4kA-0%/﹢30% @2kV

8kA-0%/﹢30% @4kV

12kA-0%/﹢30% @6kV

16kA-0%/﹢30% @8kV

20kA-0%/﹢30% @10kV

- General data: 

Power supply: AC220V ± 10%,50/60Hz

Temperature: -10~+40℃

Humidity: 35%-85%RH(no condensation)

Ground resistance: No more than 0.5Ω

Dimension: 1730x600x700mm(35U)

Weight: Approx. 200kg

No conductive dust, no fire or explosion danger, no corrosive metal and insulation air.

Waveform of mains voltage is sine wave. The THD is <5%.

Datasheet

혜택 소식 받아보기

대량 할인, 도매 가격 업데이트 및 신제품 소식을 이메일로 받아보세요.

구독하면 당사의 서비스 이용약관개인정보 처리방침에 동의하는 것으로 간주됩니다.

빠른 지원

인증된 전문가에게 직접 연결