Analytical Industries GPR-1600 산업용 가스용 산소 분석기 (0-10, 0-100, 0-1000 ppm, 0-1%)
인증된 전문가에게 직접 연결
인증된 전문가에게 직접 연결
Measurement range: 0-10, 0-100, 0-1000 ppm, 0-1% (0-25% calibration only)
Accuracy: < ±2% of selected range; ±0.1% at constant conditions after calibration with 95-100% oxygen (GPR-3100)
Response time: T90 < 10 seconds
Recovery time: 60 sec in air to < 10 ppm in < 1 hour on N2 purge
Sensitivity (LDL): 50 ppb
Linearity: < 0.5% of scale
Sensor model: GPR-12-333; XLT-12-333 for gas mixture with > 0.5% CO2
Sensor life at 25°C and 1 atm: 24 months in < 1000 ppm O2
Calibration interval: Typically: 1-3 months
Inlet pressure: 0.34 – 2 barg (5-30 psig) with atmospheric vent
Flow rate (constant): 0.5 - 1.0 Nl/min (1-2 SCFH)
Gas connections: 1/4" compression tube fittings
Display Graphical: LCD 12.7 x 7cm (5 x 2.75”); resolution 0.01
Enclosure: Painted aluminum; See individual mounting options for dimensions
Compensation: Barometric pressure and temperature; Temperature controlled sample system and sensor (GPR-3100 only)
Analog output: 4-20 mA isolated, 0-1V, and 0-5V
Range ID: 1-5 V or 4-20mA, optional relay contacts
Communications: Choose from RS485, RS232 or USB
Alarms: Two user-adjustable alarm relays
Operating temperature:
GPR sensor: 5°C to 45ºC (41°F to 113°F)
XLT sensor: -10°C to 45°C (14°F to 113°F)
Power Universal: 100-240 V AC