Anton Paar Xsample 631 Heated Sample Changer (56 samples; 95 °C / 200 °F)

문의하기
안전한 결제
품질 보장
간편 교환 및 반품
배송 가능

Maximum temperature: 95 °C (203 °F) at 20 °C (68 °F) ambient temperature, no cooling

Viscosity range with DMA M

Operation with internal air pump: Up to 10,000 mPa·s at 95 °C heating and measuring temperature

Operation with 2 bar external compressed air: Up to 36,000 mPa·s at 95 °C heating and measuring temperature

Viscosity range with SVM x001: 0.2 mPa·s to 5,000 mPa·s at 95 °C heating and measuring temperature

Filling mode: Pressure mode:

with integrated piston pump

with air pump of the master instrument

with compressed air

Minimum sample volume per measurement: Approx. 4.5 mL with DMA M

Approx. 6 mL with SVM X001

Number of sample vials

Magazine, heated positions only: 36 positions for 12 mL vials (only glass vials allowed!)

Xsample heated / unheated positions: 32 positions heated / 24 positions unheated for 12 mL vials (only glass vials allowed!)

Automatic filling and drying: Internal air pump of the master instrument or compressed air selectable

Compressed air supply (optional): Max. 2.0 bar (29 psi) relative

Quality of air necessary: Class 3 according to ISO 8573-1:

max. particle size: 5 µm

max. pressure dew point: -20 °C (-4 °F)

max. oil content: 1 mg/m3

Environmental conditions (EN 61010): Indoor use only

Ambient temperature: 15 °C to 35 °C (59 °F to 95 °F)

Air humidity: 10 % to 90 % relative humidity, non-condensing

Pollution degree: 2

Overvoltage category: II

Altitude: Max. 3,000 m (9,840 ft)

Max. temperature of heated parts in case of malfunction: 160 °C (320 °F)

Power supply: Power is supplied by the master instrument and the Xsample heating controller