Range:
H2: 0~100%
He: 0~2%
Ar: 0~10%
CO2: 0~20%
CH4: 0~15%
H2 in Blast Furnace Gas: 0~10%
H2 in acid converter gas: 0~20%
H2 in Wood Gasification: 0~30%
Detection principle: Thermal Conductivity Sensor
Accuracy: ±2%FS
Repeatability: ±1%FS
Stability: ±1%FS/7d
Response time: T90<30s
Charging power supply: 85~264VAC,50/60Hz
Power consumption of analyzer: Less than 15VA
Communication: RS485(Standard Matching)/RS232(Optional Matching)
Alarm: Two sets of switching signals (one set of state alarm and one set of concentration alarm) Relay Contact Capacity: 24V DC, 0.2A
Analog output signal: Group 1,4-20mA/0-20mA/0-1V/0-5V/0-10V/1-5V,freely set mode Load:less than 500Ω
Environmental temperature: -10~45°C Environmental humidity: less than 80% RH
External control: Quantity: Group 1. Control signal:9-28VDC. Purpose: Controlling pumps(selection)
Sample pressure: atmospheric pressure ±10%(outlet should be atmospheric pressure)
Analyzer weight: net weight 4.25 kg