인증된 전문가에게 직접 연결
인증된 전문가에게 직접 연결
Range: 0~10000ppm H2S
Resolving power: 0.1ppm
Accuracy: ±2%FS
Repeatability: ±1%FS
Stability: ±1%FS/7d
Response time: T90<35s(0~20ppm)
Recovery time: <60s
Protection level: IP64
Measurements: On-line
Calibration cycle: One year (recommendation)
Measuring instrument weight: Net weight 5.0 kg,about 15 kg after packaging
Power supply: 24V DC
Power waste: Less than 15VA
Communication: RS485(Standard Matching)/RS232 (Optional Matching)
Alarm: Alarm point:1; Relay Contact Capacity:24VDC,0.2A
Analog output signal: Group 1,4-20mA/0-20mA/0-1V/0-5V/0-10V/1-5V,output mode can be set freely; Load:less than 500Ω
Environmental temperature: -10~45°C; Environmental humidity:Less than 90% RH
Sample flow rate: 400-500ml/min(recommended 450ml/min)
Sample pressure: Atmospheric pressure ±10% (outlet should be atmospheric pressure)
Inlet and outlet: 1/4-inch NPT internal thread,stainless steel material (standard 1/4-6 card-plastic joint)
Sensor life: More than 2 years (normal use)
Instrument life: More than 5 years (normal use)
Cabinet size: L x W x D= 321 mm x 200 mm x 134.5 mm
Installation mode: Wall hanging type
Installation dimensions: L*W=335 mm*180 mm
L*W*H=300*200*175mm; Flowmeter:Selection