Electrical Specification
- Channels: 1
- Current channel
+ Input connection: BNC
+ Voltage range: 0 - 1 VAc
+ Resolution: 0.1 mA
+ Accuracy (50/60 Hz)0 to 10 mV: unspecified ; 10 to 50 mV: ±(0.5 % R + 1 mV); 50 to 1000 mV: ±(0.5 % R + 0.5 mV)
- Voltage channel
+ Input connection: 2 recessed banana jacks
+ Voltage range: 0 - 600 VAc
+ Resolution: 0.1 V
+ Accuracy (50/60 Hz): 0 to 5 V: unspecified ; 5 to 50 V: ±(0.5 % R + 1 V); 50 to 600 V: ±(0.5 % R + 0.5 V)
- Sample rate: 64 samples/cycle
- Storage rate: Programmable from 125 ms to 1 day
- Storage modes: Stop when full, FIFO and Extended Recording Mode (XRM™)
- Recording length: 15 minutes to 8 weeks, programmable using DataView®
- Memory: 240,000 measurement (512 kB). The recorded data is stored in non-volatile memory & retained even if the battery is low or removed
- Communication: USB 2.0 optically isolated
- Power source: 2 x 1.5 V AA-cell alkaline batteries
- Battery life: 100 hours to > 45 days (depending on storage rate/recording length)
Mechanical Specification
- Dimensions: 136 x 70 x 32 mm (5.38 x 2.75 x 1.28")
- Max conductor size: Depends on current probe
- Weight (with battery): 180 g (6.4 oz)
- Casing: UL94-V0
- Vibration: IEC 60068-2-6 (1.5 mm, 10 to 55 Hz)
- Shock: IEC 60068-2-27 (30G)
- Drop:  IEC 60068-2-32 (1 m)
Environment Specification
- Operating temperature: -10 to +50 °C
- Storage temperature: -20 to +60 °C

State at delivery:  type A to 5-pin mini-B USB 2 m, DataView® CD, two 1.5 V AA-cell alkaline batteries and user manual.

datasheet
user guider

Advantages

  • Single-phase voltage and current logger
  • Apparent power
  • Diagnostics on residential, commercial and industrial networks
  • Own power supply

Description

The Simple Logger II L562 is a logger which records voltage and current simultaneously. It can be used to observe a trend recording of the apparent power on the installation.

The new Simple Logger II range of data loggers from Chauvin Arnoux offers an excellent quality/price ratio when their specifications and functions are taken into account. Their compact design and their battery power supply mean they can be set up anywhere, even in confined spaces.


Specifications:

  • One voltage input (banana plugs) from 0 V to 600 VAC
  • One current input (BNC)
  • 240,000 measurements
  • Recording for up to 8 weeks
  • USB port
  • IEC 61010 300V CAT IV and 600 CAT III