인증된 전문가에게 직접 연결
인증된 전문가에게 직접 연결
- Measurement range:
+J : -210 to + 1,200 °C
+K : -200 to + 1,372 °C
+T : -250 to + 400 °C
+N : -200 to + 1,300 °C
+E : -150 to + 950 °C
+R: 0 to + 1,767 °C
+S: 0 to + 1,767 °C
- Resolution : Display in °C: ø < 1,000 °C : 0.1 °C and ø ≥ 1,000°C : 1 °C
- Intrinsic uncertainty in °C :
+ J. K. T. N. E: ø ≤ - 100°C ± (0.2% Reading + 0.6°C) ; - 100°C < ø ≤ + 100°C ± (0.15% R + 0.6°C) ; + 100°C < ø ± (0.1% R + 0.6°C)
+ R. S: ø ≤ + 100°C ± (0.15% R + 1.0°C) ; + 100°C < ø ± (0.1% R + 1.0°C)
FUNCTIONS
- Recording :
+ Manual start/stop on the product : Short press on MEM: spot recording / Long press on REC: recording at the rate of the current mode
+ Programmed recording: Start date, recording rate (interval from 1 minute to 2 hours) and end date can be customized with the PC software
- Alarms: Alarm thresholds set using the software. Visual alert on the product in the event of an overrun. Recording can be triggered on alarm thresholds
- Data storage: More than 1 million points
- Min-Max-Hold: Yes
- Differential measurement : yes
- Backlighting:Yes
- Automatic power off:Yes
POWER SUPPLY
- Type: 3 x 1.5V AA / LR6 alkaline batteries. Mains connection possible with the mains / micro USB adapter proposed as an accessory
- Battery life: 1,000 hrs (portable mode)
PHYSICAL SPECIFICATIONS
- Interfaces: 2 possible communication modes: Bluetooth and USB
- Fixations : Casing equipped with a magnet, a wall-mount system and a slot for suspension of the product. Compatible with the Multifix accessory. Sheath available as an accessory
- Connections: compensated miniature female connectors
- Dimensions / weight : 150 x 72 x 32 mm / 260 g with batteries
- Ingress protection : IP54 casing
- Product operating range: Temperature from -10 to +60 ° C / Humidity from 10 to 90 %RH
- Standards: IEC61010-1 / IEC 61326-1
GENERAL SPECIFICATIONS
Windows Data Logger Transfer software functions: Real-time mode. Automatic report generation in Word format. Data format compatible with the Dataview software. Graph or value-table presentation . Data export in graph or Excel table format
State at delivery: Bag, 3 x 1.5 V AA alkaline batteries, USB cable, test report and quick start-up guide
Additional specifications:
- 2 inputs for K, J, T, N, E, R or S thermocouples.
- Measurement range: - 210 °C to +1,767 °C
- Intrinsic uncertainty:
(J, K, T, N, E) θ ≤ - 100 °C : ± (0.2% R + 0.6 °C)
(J, K, T, N, E) -100 °C < θ ≤ + 100 °C : ± (0.15% R + 0.6 °C)
(J, K, T, N, E) +100 °C < θ: ± (0.1% R + 0.6 °C)
(R, S) θ ≤ + 100 °C : ± (0.15% R + 1.0 °C)
(R, S) +100 °C < θ: ± (0.1% R + 1.0 °C)
- Resolution: 0.1 °C or 1 °C
- Min, Max, Hold and ΔT functions, Alarms / °C or °F / Backlighting
- Recording up to 1 million points
- USB or Bluetooth interfaces
- IP54 casing
- Miniature female compensated connector
- Dimensions: 150 x 72 x 32 mm
- Weight: 260 g with batteries
- Compatible with the Multifix accessory
- Shockproof protective sheath available as an accessory