Chromatotec A11000 VOC Analyzer (20 mL/min; 0.5 ~ 100 ppb)

문의하기
안전한 결제
품질 보장
간편 교환 및 반품
배송 가능

airmoVOC C2-C6 (4U)

Light VOC Analysis: C2–C6 (Alkanes, Halogenated Compounds, Light Terpenes)


Detection Limit

10 ppt for Butane

15 ppt for 1.3-Butadiene


Detection Range: 0.5 to 100 ppb / amplification3 in standard


Relative Standard Deviation

< 0.3 % over 48h (RT)

< 2 % over 48h (Conc.)


Results

Data storage (SSD)

4–20 mA current output (option)

MODBUS / JBUS or MGS1 communication protocol (option)


Cycle Time: 30 min in standard (60 min in option)

Gas Supply:

H2 (FID and carrier gas): 30 mL/min (inlet 2 bars; 1/16’’ Swagelok)

Air (FID): 180 mL/min (inlet 3 bars; 1/8’’ Swagelok)

Sample inlet (vacuum pump): 1/4’’ Swagelok

Pneumatic valve: 90 mL / commutation


Sample Volume: 20 to 240 mL (programmable) (Sampling time up to 40 minutes)


Sampling Inlet:

Sample by-pass: 20 mL/min


Power Supply

Main: 230 V / 115 V or 50 Hz / 60 Hz

Battery: 24 V (option)


Electrical Consumption

Mean: 150 VA, Peak: 360 VA


Dimensions and Weight

Rack: 19’’ (4U)

Height: 180 mm

Width: 482 mm