Detection limit

Loop

Benzene & Toluene: ~ 1 ppb

Ethylbenzene & m+p-Xylenes: ~ 2 ppb (with default settings) / o-Xylene: ~ 4 ppb


Configuration & cycle time

Loop

BTEX: 10 min

Benzene + 1.3 butadiene: from 10 to 15 min

Benzene: 5 min (3 min in option)

Other application on demand


Detection range

Loop

0-1000 ppb

0-10 ppm / 0-100 ppm / 0-1000 ppm


Measurement

Detector: PID

Temporal resolution: 0.1 seconds

Response time: One measurement every 10 minutes (default settings)

Analysis sample condition: Gas T°: 5 - 40°C; Gas RH: 20 - 90%; Atmospheric pressure

Calibration: Gaseous BTEX mixture or gaseous Benzene


Sampling

200μL loop or carbon trap

Gas flow rate: between 10 to 100 mL min-1

Carrier gas: Nitrogen 4 bar inlet pressure and 2.5 mL min-1

Supply inlet connection: 1/8’’


Instrument supply

Power supply: Input 100 - 240V ±10%; 1.5 A max; 47 - 63 Hz - Output 15V; 6.67A 100W

Autonomy on battery: Up to 4h

Power consumption: max 75 W


General

Dimensions (analyzer): 32×28×15 cm; 6.0 kg

Dimensions (suitcase): 56 x 45 x 25 cm; 18 kg

Operational conditions: 0 - 40°C / 20 - 80% RH

Storage conditions: -20°C - +40°C / 0 - 85% RH

Display: 7” TFT display; resolution 800 x 480; integrated touchscreen

Autonomy: BTEX version: 21 days for Gas - 4 hours for battery. Can be reduced for different applications