For full functionality of this site it is necessary to enable JavaScript.
EMIN.VN
0
Product image

Cosmopia ER-105EXMT Walk-In Type Temperature Chamber (-40 ~ 80℃; 9.7m²)

문의하기
안전한 결제
품질 보장
간편 교환 및 반품
배송 가능

Walk-in type temperature and humidity chambers: Water-cooling

Temperature Range: -40 to 80 ℃

JTM K07

Temp./Humidity Fluctuation: ±0.3 ℃

Temp./Humidity Gradient: ±2.5 ℃

Spatial Temp./Humidity Deviation: ±2.0 ℃

Temp. Change Rate (Rise): 2.3°C/min. (-28 to 68°C)

Temp. Change Rate (Drop): 0.6°C/min. (68 to -28°C)

Time to Reach Extreme Temp. (Rise): Within 30 min. (20 to 80°C)

Time to Reach Extreme Temp. (Drop): Within 120 min. (20 to -40°C)


Testing Room Floor Space: 9.7 m²

External Dimensions (W×D×H): 3,600 × 2,700 × 2,325 mm

Internal Dimensions (W×D×H): 3,450 × 2,550 × 2,100 mm

Exterior / Interior Materials: Color steel plate (ivory) / Stainless steel plate

Floor Load Capacity: 5.9 kN/m²

Door (Width x Height): 830 x 1,800 mm (Single opening, 1 location)

Observation Window (W×H): 190 x 320 mm (Door area, 1 location)

Inner Light: 2 lamps (60 W capacity)

Cable Hole: ø50 mm, 1 location (with rubber plug)

Temperature Control Unit: EU-125EXMT

Defrosting Method: Off-cycle defrosting (refrigerator operation stopped, room temperature 5 to 40°C), temperature-increase defrosting (room temperature -40/-30/-10 to 5°C)

Power Supply: Three-phase 200V 50/60Hz

Maximum Load Current: 167 A

ELB Capacity: 200 A

Coolant Water Volume: 8400 L/h (at 32℃ inlet temperature)

Coolant Water Pressure: 0.1 to 0.5 MPa

Coolant Water Temperature: 18 to 32 ℃

Coolant Piping Dimensions (Inlet/Outlet): Rc2 / Rc2

[High load mode] Heat generation load (40°C): 10 kW

[Standard mode] Heat generation load (40°C): 6 kW

Datasheet