Dansensor CheckPoint 4 Portable Headspace Analyzer (O₂, CO₂: 0–100%)
O₂ (optical sensor)
Sample volume (minimum): 5 ml
Sample time (minimum): 7 sec
Measuring range: 0-100%
Resolution: 0.01% oxygen
Accuracy: ±(0.25% O₂ + 2% of readout)
Applications: Most food applications, except those that utilize alcohol (i.e. fresh bread)
O₂ (electrochemical sensor)
Sample volume (minimum): 5 ml
Sample time (minimum): 7 sec
Measuring range: 0-100%
Resolution: 0.01% oxygen
Accuracy: ±(0.25% O₂ + 2% of readout)
Applications: All food applications
CO₂ (infrared single beam sensor)
Sample volume (minimum): 5 ml
Sample time (minimum): 7 sec
Measuring range: 0-100%
Resolution: 0.01% carbon dioxide
Accuracy:
±2% CO₂ in range 0-20%
±3% CO₂ in range 20-100%
Applications: Most food applications with carbon dioxide
Data export: WiFi (WPA, WPA 2, WPA enterprise)
Dimensions & weight: 77 x 100 x 175 mm (HxWxD), 0.7 kg
Ambient temperature: Operational: 0 to +40°C
Relative humidity: Operational: 10 to 90 %RH (non-condensing) Storage: less than 95 %RH (non-condensing)
Power supply: Battery powered; up to 2000 measurements on a fully charged battery at 20 °C; recharge time maximum 5 hours
Consumables and accessories: Standard consumable kit (1 sampling kit, 1000 septa, 10 needles (Ø 0.8) & 10 water trap filters), metal needles, can piercer for testing cans and wine bottles