Power Supply: DC 7.4V 2600mAh rechargeable lithium battery, the voltage of full charged battery is 8.4V.
Display Mode: 5 inch TFT true-color touch screen
Rated Voltage: AC 220V (The rated voltage of tester is AC 220V during leakage protection switch testing)
Test Line Voltage: 220V/380V
Leakage Current Gear: 5mA, 10mA, 15mA, 20mA, 30mA, 40mA, 50mA, 60mA, 100mA, 150mA, 200mA, 250mA, 300mA, 500mA, 600mA, 1000mA
Gear Current Accuracy: ±10%rdg±5dgt
Test Range Of Leakage Current Operating Time: 0~999ms
Accuracy Of Leakage Operating Time: ±5%rdg±3dgt
AC Voltage Measurement Range: 0~600V
AC Voltage Accuracy: ±2%rdg±3dgt
Test Mode
Manual: X1 mode: 10mA, 20mA, 30mA, 100mA, 300mA, 500mA
X1/2 mode: 5mA, 10mA, 15mA, 50mA, 150mA, 250mA
X2 mode: 20mA, 40mA, 60mA, 200mA, 600mA, 1000mA
Automatic:Auto (5mA~1000mA)
Failure Message Indication: LCD displays “999” timeout indication, indicates that tested rated operating current gear is low or leakage protector has fault.
No. Of Memory Groups: 100 groups
Length of test line: 1.5m
Test Mode: Manual or automatic
Tester Size: 215mm×178mm×83mm
Weight: Host: about 944g (include battery); total weight: about 2kg(include accessory)
Battery Voltage: When the battery voltage drops to 7.0V, the low battery voltage prompts the tester need to be charged in time.
Working Current: 365mA Max
Insulation Withstand Voltage: 100MΩ or greater, AC2kV/rms(between the housing and screw)
Suitable Safety Standard: IEC1010-1, IEC1010-2-032, Pollution level 2, CAT Ⅲ(600V) IEC61326(EMC Standard)
Working Temperature & Humidity: -10℃~40℃; below 80%RH
Store Temperature & Humidity: -10℃~60℃; below 70%RH