Mains Voltage Measurement

Range: 90.0 to 264.0 V ac rms

Accuracy: ±(2 % of reading + 0.2 V)

Earth Resistance

Modes: Two wire

Test Current: >200 mA dc

Range: 0.000 Ω to 20.000 Ω

Accuracy: ± (1 % of reading + 0.010 Ω)

Resistance tests: Earth resistance and point to point

Equipment Current

Mode: AC RMS

Range: 0.0 to 20.0 A

Accuracy: ±(5 % of reading + (2 counts or 0.2 A, whichever is greater))

Duty cycle

15 A to 20 A, 5 min. on/5 min. off

10 A to 15 A, 7 min. on/3 min. off

0 A to 10 A, continuous

Leakage Current

Modes: True-rms

Patient Load Selection:

AAMI ES1-1993 Fig. 1

IEC 60601: Fig. 15

Crest factor: less than or equal to 3

Ranges: 0.0µA to 1999.9µA Accuracy

DC to 1 kHz: ± (1 % of reading + (1 µA , whichever is greater))

1 to 100 kHz: ± (2.5 % of reading + (1 µA, whichever is greater))

100 kHz to 1 MHz: ± (5 % of reading + (1 µA, whichever is greater))

Leakage Tests

Ground wire (earth)

Chassis (enclosure)

Direct equipment

Point to point

General Specifications

Display: LCD display

Modes of Operation: Manual

Temperature

Operating: 0 °C to 50 °C (32 °F to 122 °F)

Storage: -20 °C to 60 °C (-4 °F to 140 °F)

Humidity: 10 % to 90 % non-condensing

Altitude

120 V ac mains supply voltage: 5000 m

230 V ac mains supply voltage: 2000 m

Power ratings

115 Volt power outlet

90 to 132 V ac rms, 47 to 63 Hz, 20 A maximum

230 Volt power outlet

180 to 264 V ac rms, 47 to 63 Hz, 16 A maximum

Power input: 115V 20A - 2.6kVA and 230V at 16A - 4.2kVA

Physical case

Weight: 0.7 kg (1.5 lb)

Dimensions (WxDxH): 20 cm x 18 cm x 5 cm (7.9 in x 7.1 in x 2.1 in) 