Flow and Volume

Mean flow range: 0 – 3000 mL/h (steady flow), 0 – 1500 mL/h (peristaltic flow)

Mean flow resolution: 0.001 mL/h

Mean flow accuracy (after > 0.1 mL and > 10 s): ± (1% + 0.005 mL/h) at < 500 mL/h, otherwise ± 2%

Volume range: 0 – 100,000 mL

Volume resolution: 0.001 mL

Volume accuracy (after > 0.1 mL and > 10 s): ± (1% + 0.003 mL) at < 500 mL/h, otherwise ± 2%

Time range: 0 s – 1000 h

Time accuracy: ± (0.2% + 0.2 s)

Time trigger volume: ≤ 3 μL

Graphs: Instant flow and mean flow with up to 1 s resolution (gradually reduced after 1 hour), IEC 60601-2-24 trumpet curve

Occlusion

Peak pressure range: -200 – +2600 mmHg (-3.8 – +50 psi)

Peak pressure resolution: 1 mmHg (0.01 psi)

Peak pressure accuracy: ± (1% + 5 mmHg (0.1 psi)), 50 ms moving average

Time to peak range: 0 s – 1000 h

Time to peak accuracy: ± (0.2% + 0.05 s)

Time to alarm: 0 s – 1000 h

Time to alarm accuracy: Operator dependent, manual stop button

Residual volume accuracy: ± (2% + 0.01 mL)

Graphs: Pressure with up to 50 ms resolution (gradually reduced after 3 minutes)

Bolus and PCA

Bolus volume range: 0.01 – 100,000 mL

Bolus volume accuracy (after > 1 s): ± (2% + 0.01 mL)

Bolus flow range: 10 × base flow – 3000 mL/h (steady flow), 10 × base flow – 1500 mL/h (peristaltic flow)

Bolus flow accuracy (after > 0.1 mL and > 10 s): ± (1% + 0.005 mL/h) at < 500 mL/h, otherwise ± 2%

Base flow range: 1 – 100 mL/h

Base flow accuracy (after > 0.1 mL): ± (1% + 0.005 mL/h)

Bolus duration range: 0 s – 1000 h

Bolus duration accuracy: ± (0.2% + 0.2 s)

Graphs: Instant flow with up to 1 s resolution (gradually reduced after 1 hour), numbered boluses

Back Pressure

Range: -200 – +600 mmHg (-3.8 – +11.6 psi)

Additional flow uncertainty: ± ΔP × 0.001 mL/h

Additional volume uncertainty: ± (test duration in hours) × ΔP × 0.001 mL

Where ΔP is the average pressure difference between inlet and outlet, in mmHg.

Measurement Units

Pressure: mmHg, kPa, psi, bar

Volume: mL, g (1 mL = 0.998 g)

General

Safety standard: Complies with IEC 61010-1:2010, pollution degree 2

Infusion device standard: For testing according to IEC 60601-2-24:1998

EMC standard: Complies with IEC 61326-1:2012

Dimensions (W × D × H): 95 mm × 210 mm × 260 mm (11.6 in × 8.3 in × 10.1 in)

Weight: 3.8 kg (8.4 pounds)

Operating temperature: +15 – +30°C (+59 – +86°F)

Storage temperature: -20 – +70°C (-4 – +158°F), when drained

Atmospheric pressure: 70 – 107 kPa, altitude up to 3000 m (10,000 ft)

Humidity: < 90% relative humidity, non-condensing

Power: 19 V DC, 3 A

Connectivity: 1 × USB-C for PC communication, 3 × USB-A for peripherals

Display: 10-inch multi-touch, 800 × 600 pixels

Data storage: > 10,000 measurements

Channels: 4