Range: –20 °C to 150 °C
Stability:
±0.005°C at –20°C (ethanol)
±0.005°C at 25°C (water)
±0.007°C at 150°C (oil 5012)
Uniformity:
±0.007 °C at –20 °C (ethanol)
±0.00 7 °C at 2 5°C (water)
±0.010°C at 150 °C (oil 5012)
Heating Time: 120 minutes, from 25 °C to 150 °C (oil 5012)
Cooling Time: 110 minutes, from 25 °C to –20 °C (ethanol)
Stabilization Time: 15–20 minutes
Temperature Setting: Digital display with push-button data entry
Set-Point Resolution: 0.01°; 0.00018° in high-resolution mode
Display Resolution: 0.01 °
Digital Setting Accuracy: ±1°C
Digital Setting Repeatability: ± 0.01 °C
Access Opening: 120 x 172 mm (4.7 x 6.8 in)
Depth:
457 mm (18 in) without Liquid-in-Glass Thermometer Cal Kit
482 mm (19 in) with Liquid-in-Glass Thermometer Cal Kit
Wetted Parts: 304 stainless steel
Power: 115 VAC (±10 %), 60 Hz, 14 A or 230 VAC (±10 %), 50 Hz, 7 A, specify
Volume: 15,9 Liters
Size (WxDxH): 1067 x 356 x 788 mm (940 mm from floor to tank access opening)
Weight: 62 kg (137 lb.
Datasheet
manual
| Specifications |
| Range |
| 6331 |
35 °C to 300 °C |
| 7321 |
–20 °C to 150 °C |
| 7341 |
–45 °C to 150 °C |
| 7381 |
–80 °C to 110 °C |
|
| Stability |
| 6331 |
±0.007 °C at 100 °C (oil 5012) ±0.010 °C at 200 °C (oil 5017) ±0.015 °C at 300 °C (oil 5017) |
| 7321 |
±0.005°C at –20°C (ethanol) ±0.005°C at 25°C (water) ±0.007°C at 150°C (oil 5012) |
| 7341 |
±0.005°C at –45°C (ethanol) ±0.005°C at 25°C (water) ±0.007°C at 150°C (oil 5012) |
| 7381 |
±0.006°C at –80°C (ethanol) ±0.005°C at 0°C (ethanol) ±0.005°C at 100°C (oil 5012) |
|
| Uniformity |
| 6331 |
±0.007 °C at 100 °C (oil 5012) ±0.017 °C at 200 °C (oil 5017) ±0.025 °C at 300 °C (oil 5017) |
| 7321 |
±0.007 °C at –20 °C (ethanol) ±0.00 7 °C at 2 5°C (water) ±0.010°C at 150 °C (oil 5012) |
| 7341 |
±0.007 °C at –45 °C (ethanol) ±0.007 °C at 25 °C (water) ±0.010 °C at 150 °C (oil 5012) |
| 7381 |
±0.007 °C at –80 °C (ethanol) ±0.007 °C at 0 °C (ethanol) ±0.007 °C at 100 °C (oil 5012) |
|
| Heating Time† |
| 6331 |
130 minutes, from 40 °C to 300 °C (oil 5017) |
| 7321 |
120 minutes, from 25 °C to 150 °C (oil 5012) |
| 7341 |
120 minutes, from 25 °C to 150 °C (oil 5012) |
| 7381 |
60 minutes, from 25 °C to 100 °C (oil 5012) |
|
| Cooling Time† |
| 6331 |
14 hours, from 300 °C to 100 °C (oil 5017) |
| 7321 |
110 minutes, from 25 °C to –20 °C (ethanol) |
| 7341 |
130 minutes, from 25°C to –45°C (ethanol) |
| 7381 |
210 minutes, from 25 °C to –80 °C (ethanol) |
|
| Stabilization Time |
|
| Temperature Setting |
| |
Digital display with push-button data entry |
|
| Set-Point Resolution |
| |
0.01°; 0.00018° in high-resolution mode |
|
| Display Resolution |
|
| Digital Setting Accuracy |
|
| Digital Setting Repeatability |
|
| Access Opening |
| |
120 x 172 mm (4.7 x 6.8 in) |
|
| Depth |
| |
457 mm (18 in) without Liquid-in-Glass Thermometer Cal Kit |
| |
482 mm (19 in) with Liquid-in-Glass Thermometer Cal Kit |
|
| Wetted Parts |
|
| Power† |
| 6331 |
115 VAC (±10 %), 50/60 Hz, 14.8 A or 230 VAC (±10 %), 50/60 Hz, 7.4 A, specify |
| 7321 |
115 VAC (±10 %), 60 Hz, 14 A or 230 VAC (±10 %), 50 Hz, 7 A, specify |
| 7341 |
115 VAC (±10 %), 60 Hz, 16 A or 230 VAC (±10 %), 50 Hz, 8 A, specify |
| 7381 |
230 VAC (±10 %), 50 or 60 Hz, specify, 10 A |
|
| Volume |
|
| Size (HxWxD) |
| |
1067 x 356 x 788 mm (940 mm from floor to tank access opening) [42 x 14 x 31 in (37 in from floor to tank access opening)] |
|
| Weight |
| 6331 |
41 kg (90 lb.) |
| 7321 |
62 kg (137 lb.) |
| 7341 |
68 kg (150 lb.) |
| 7381 |
91 kg (200 lb.) |
|
| Automation Package |
| |
Interface-it software and RS-232 included (IEEE-488 optional) |
|
| |
| |
†Rated at nominal 115 V (or optional 230 V) |
|