Flow Measurement
Flow ranges: ±3 cc/min, ±6 cc/min, ±30 cc/min, ±300 cc/min
Flow Units:
±3.000 cc/min, ±180.0 cc/h, ±50.00 mm3/s, ±3.000 ml/min, ±180.0 ml/h
±6.000 cc/min, ±360.0 cc/h, ±100.0 mm3/s, ±6.000 ml/min, ±360.0 ml/h
±30.00 cc/min, ±1800 cc/h, ±500.0 mm3/s, ±30.00 ml/min, ±1800 ml/h
±300.0 cc/min, ±18000 cc/h, ±5000 mm3/s, ±300.0 ml/min, ±18000 ml/h
Accuracy @ 20°C: <1% R + 0.003 cc/min, <1% R + 0.006 cc/min, <1% R + 0.03 cc/min, <1% R + 0.3 cc/min
Temperature Coefficients:
Zero: Semi-automatic Zero facility
Span: < 0.1% per °C
Display: 8 character LCD
Long Term Drift (span): < 1% per year
Electrical
Power Requirements: 9 volt battery (PP3/6LR61/MN1604)
Battery Life: Up to 100 hours
Operating Temperature: 0 to 50°C
Pneumatic
Media Compatibility: Clean dry air or non corrosive gas. Viscosity corrected for air only.
Overload: 20 x Range
Pneumatic Connection: ⅛” BSP female input (option for Quick Connector)
Vent port: M5 (DO NOT RESTRICT)
Static Pressure: Vacuum to 14 bar gauge
Effective Volume: 0.8cc (input port up to control valve)
Construction
Enclosure: ABS Case
Dimensions: 222 x 205 x 79mm (W x H x D)
Media Compatibility: Clean dry non-corrosive gas
Weight: 1.2kg