Gastec 101L 빠른 측정 감지기 튜브 가솔린 CnHm (30 ~ 2000ppm)
Model101L
문의하기
안전한 결제
품질 보장
간편 교환 및 반품
배송 가능
Indicator: 30 to 1000 ppm
Measuring Range:
+ 30 to 1000 ppm, Number of Pump Strokes 2
+ 1000 to 2000 ppm, Number of Pump Strokes 1
Sampling Time: 2 minute per pump stroke
Colour Change: Brownish yellow → Dark green
Reaction Principle: CnHm + Cr6++ H2SO4 → Cr3+
Coefficient of Variation: 10% (for 30 to 300 ppm), 5% (for 300 to 1000 ppm)
Nr. of tests: 10
Pump: GV-100S/110S
Performance
| Measuring Range |
|
||
| Number of Pump Strokes |
|
||
| Correction Factor |
|
||
| Sampling Time | 2 minutes per pump stroke | ||
| Detecting Limit | 5 ppm (n=2) | ||
| Colour Change | Brownish yellow → Dark green | ||
| Reaction Principle | CnHm + Cr6++ H2SO4 → Cr3+ | ||
| Coefficient of Variation | 10% (for 30 to 300 ppm), 5% (for 300 to 1000 ppm) | ||
| Shelf Life | 3 Years | ||
| Corrections for temperature & humidity | Unnecessary | ||
| Store the tubes at cool and dark place. | |||
Possible coexisting substances and their interferences (NOTE)
| Substance | Concentration | Interference | Change colour by itself to |
| Hydrogen sulphide | + | Dark brown | |
| Acetylene | + | Dark brown | |
| Aromatic HCs | + | Dark green | |
| Esters, Ethers,Alcohols | + | Dark green | |
| sulphur dioxide | + | Dark green | |
Organic solvents ( C3) |
+ | Dark green |
Other substance measurable with this detector tube
| Substance | Correction Factor | No. of pump strokes | Measuring Range |
| Allyl chloride | Factor: 34 | 1/2 | 0.1 to 3.4% |
| Heptane | Factor: 1.0 | 1 or 2 | 30 to 2000 ppm |
| Isobutene | Factor: 22 | 1 | 0.07 to 2.2% |
| Calibration gas generation | Static gas dilution method |
| TLV-TWA | TLV-STEL | Explosive range |
| 300 ppm | 500 ppm | 1.4 to 7.6% |
빠른 지원
인증된 전문가에게 직접 연결


C3)