For full functionality of this site it is necessary to enable JavaScript.
EMIN.VN
0
Product image

Gastec 132D 수동형 Dosl 튜브 Trichloroethylene Cl2C:CHCl 검지관 (3 ~ 300 ppm)

문의하기
안전한 결제
품질 보장
간편 교환 및 반품
배송 가능

Indicator: 25 ~ 300 ppm
Measuring Range: 3 ~ 300 ppm
Sampling Hours: 1 to 8 hours
Colour Change: Yellow → Purple
Reaction Principle: Cl2C:CHCl + PbO2 + H2SO4 → HCl, HCl + Base → Chloride 
Coefficient of Variation: 10% (for 25 to 300 ppm-hr)
Nr. of tests: 10

Catalog

Manual

Performance
Measuring Range 3 to 300 ppm
Sampling Hours 1 to 8 hours
Colour Change Yellow → Purple
Reaction Principle Cl2C:CHCl + PbO2 + H2SO4 → HCl
HCl + Base → Chloride
Coefficient of Variation 10% (for 25 to 300 ppm-hr)
Shelf Life 1Year
Corrections for temperature & humidity Temperature correction is necessary
Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below.
Possible coexisting substances and their interferences (NOTE)
Substance Interference Change colour by itself to
Hydrogen chloride, Chlorine + Purple
1,2-Dichloroethylene, Tetrachloroethylene + Purple
Toluene, Xylene No No
Other substance measurable with this detector tube
Substance Correction Factor Sampling Time Measuring Range
trans-1,2-Dichloroethylene 2.0 1 to 8 hours 6 to 600 ppm
Chlorine 0.8 1 to 8 hours 2.4 to 240 ppm
Hydrogen chloride 0.6 1 to 8 hours 1.8 to 180 ppm
Tetrachloroethylene 0.5 1 to 8 hours 1.5 to 150 ppm
Calibration gas generation Diffusion tube method
TLV-TWA TLV-STEL Explosive range
10 ppm 25 ppm -
혜택 소식 받아보기

대량 할인, 도매 가격 업데이트 및 신제품 소식을 이메일로 받아보세요.

구독하면 당사의 서비스 이용약관 및 개인정보 처리방침에 동의하는 것으로 간주됩니다.

빠른 지원

인증된 전문가에게 직접 연결