Gastec 134L 빠른 측정 감지기 튜브 사염화탄소 CCl4 (0.25 ~ 11ppm)
Model134L
문의하기
안전한 결제
품질 보장
간편 교환 및 반품
배송 가능
Indicator: 0.25 to 5 ppm
Measuring Range:
+ 0.25 to 5 ppm, Number of Pump Strokes 2
+ 5 to 11 ppm, Number of Pump Strokes 1
Sampling Time: 2 minutes per pump stroke
Colour Change: White → Yellow
Reaction Principle:
+ CCl4 + I2O5 + H2S2O7 → COCl2,
+ COCl2 + (CH3)2NC6 H4CHO→ (CH3)2NC6 H4CHCl2 + CO2,
+ (CH3)2NC6H4CHCl2 + (C6H5)2NH(White) → Reaction product (Yellow)
Coefficient of Variation: 10% (for 0.25 to 1 ppm), 5% (for 1 to 5 ppm)
Nr. of tests: 5
Pump: GV-100S/110S
Performance
| Measuring Range |
|
||
| Number of Pump Strokes |
|
||
| Correction Factor |
|
||
| Sampling Time | 2 minutes per pump stroke | ||
| Detecting Limit | 0.04 ppm (n=2) | ||
| Colour Change | White → Yellow | ||
| Reaction Principle |
CCl4 + I2O5 + H2S2O7 → COCl2 |
||
| Coefficient of Variation | 10% (for 0.25 to 1 ppm), 5% (for 1 to 5 ppm) | ||
| Shelf Life | 1 Year | ||
| Corrections for temperature & humidity | Unnecessary | ||
| Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below. | |||
Possible coexisting substances and their interferences (NOTE)
| Substance | Concentration | Interference | Change colour by itself to |
| Hydrogen chloride | No | No | |
| Chlorine, Bromine | + | No( |
|
| Methyl bromide | + | Pale yellow ( |
|
| 1,1,1-Trichloroethane | + | No( |
Other substance measurable with this detector tube
| Substance | Correction Factor | No. of pump strokes | Measuring Range |
| Chloropicrin | 1.1 | 2 | 0.28 to 5.5 ppm |
| Calibration gas generation | Diffusion tube method |
| TLV-TWA | TLV-STEL | Explosive range |
| 5 ppm(2015) | 10 ppm(2015) | - |